80

User Setup

7.4 mA outputs

7.4.1 Outputs 0 to 3

Select the mA OUTPUTS option to set up the parameters for
all the analog outputs.

From the list available, select the mA output you wish to set.

Table 7 Analog output parameters

Description

Attribute

Choose the attribute that triggers the analog output:

Measurement channel

Temperature

Current (mV)

Note:

The attribute variable cannot be applied to Output 0, which is reserved for the

continuous live output signal.

Type

Choose either a 0-20 mA or 4-20 mA analog output

Mode

Select between linear, dual or logarithm (see graph in

Figure 20 on page 81

). This

parameter is only selectable if the attribute is set to a measurement channel.

Low

The value corresponding to the low end of the scale

Middle

The value corresponding to the mid-point of the scale (only available in dual mode)

High

The value corresponding to the high end of the scale

Advertising