7.3.1 Alarms 1 and 2

Table 3 Alarms 1 and 2 parameters

Value

Description

Mode

Limit

Trigger the alarm when the measurement is above or below a
pre-defined limit

Sample flow

Trigger the alarm when the sample flow rate is too low

Active channel

Trigger the alarm when the channel is active i.e. during the
measurement cycle

None

Alarm not required

Attributes

Channel 1

Defines the channel number on which the alarm is triggered

Limit

nnnn

Define the limit when the alarm should be triggered

Direction

Up

Trigger alarm when measurement is above the limit

Down

Trigger alarm when measurement is below the limit

Delay

nnn seconds

The delay (in seconds) before the alarm is activated

Hysteresis

nn%

Hysteresis %

Relay

N.O.

Normally open

N.C.

Normally closed

Depending on the mode set, a different screen will be
displayed.

The example left shows the Limit mode screen.

The example illustrated indicates that alarm number 2 is set
up on Channel 1, and the alarm is triggered after a delay of 15
seconds when a measurement of 21.00 ppm is exceeded.

With the mode set to either Active Channel or Sample Flow,
you only have the possibility of setting the relay, as illustrated
left.