3 input/output connections, Installation 4.4.3 input/output connections – Hach-Lange POLYMETRON 9245 User Manual User Manual
Page 43

41
Installation
4.4.3 Input/Output connections
1.
Analog output – Iout 0
is used for the current signals of the measurement. Other analog
outputs can be freely linked to different parameters like measurement, temperature – refer
to the section entitled
for details.
2.
Relays – Rel 8
is linked to the system alarm and is normally energized. Rel 7 is linked to
the warning alarm. The position of the contact represented on the diagram above
corresponds to a state when the relay is not energized. See
for more
information on the relays.
3.
Logical Inputs (2) –
one input allows the analyzer to be put on standby by providing 1 dry
circuit which will then be closed to place the analyzer on standby until that circuit is
re-opened. The second input is to launch the remote calibration (same principle as sample
measurement by-pass).
•
- In 2 (on J32) for remote calibration
•
- In 7 (on J38) to put the analyzer on standby
4.
Close the I/O box.