4 conditioning reagent, Analyzer overview – Hach-Lange POLYMETRON 9245 User Manual User Manual

Page 23

Advertising
background image

21

Analyzer Overview

3.4 Conditioning reagent

Hach Lange highly recommends the use of Diisopropylamine (DIPA) as the conditioning
reagent. Other reagents such as ammonia or ethanolamine can be applied, providing the
specification limitations imposed by amines other than DIPA are clearly understood.

The following graph (

Figure 5

below) shows the curves obtained using DIPA and ammonia as

conditioning reagents. As can be seen, the DIPA curve remains linear at a much lower sodium
concentration resulting in better accuracy, repeatability, a lower detection limit, and a lower rate
of reagent consumption.

The following table shows the comparable values of detection limit, accuracy, repeatability, and
consumption for DIPA, ammonia, and ethanolamine:

Note:

As Hach Lange recommends DIPA to be used as the conditioning reagent, all references relating to

the conditioning reagent throughout this manual will be to DIPA.

Figure 5 Calibration curves obtained with DIPA and NH

3

as reagents

DIPA

[C

6

H

15

N]

Ammonia

[NH

3

]

Ethanolamine

[H

2

N(CH

2

)

2

OH]

Lowest detection limit

0.01ppb

2ppb

5ppb

Accuracy (non-cationic applications)

The greater of

± 0.1ppb or ± 5%

of the reading

The greater of

± 1ppb or ± 5%

of the reading

The greater of

± 2ppb or ± 7%

of the reading

Accuracy (cationic applications)

The greater of

± 2ppb or ± 5%

of the reading

The greater of

± 2ppb or ± 5%

of the reading

The greater of

± 2ppb or ± 7%

of the reading

Repeatability within a 10°C variation

The greater of

< 0.02ppb or ±1.5%

of the reading

The greater of

< 0.1ppb or ±1.5%

of the reading

The greater of

< 0.2ppb or ±2%

of the reading

Consumption of 1 liter at 25°C for a
pH measurement of 10 to 10.5

approx. 13 weeks

approx. 3 weeks

approx. 7 weeks

Advertising