Analyzer Overview

3.10 Analyzer outputs

The following tables shows the outputs (screen, analog, alarm relays and RS485) generated by
the analyzer during the various processes (measurement, grab sample, calibration and sensor
reactivation).

For the grab sample and calibration processes, see

Automatic calibration cycle on page 25

,

Manual calibration cycle on page 27

, and

Grab sample measurement cycle on page 28

for a

definition of each phase referred to in the table.

The terminology used in the tables is as follows:

Frozen - output frozen at the last concentration value measured

Actual - output shows the actual concentration value being measured

For the RS485 output, several variables are accessible:

Address 0 - actual concentration value being measured

Address 2, 4, 6, 8 - concentration value for channel 1 to 4 respectively

Address 163 shows the analyzer status:

1 - number of channel

5 - grab sample mode

10 - auto calibration mode

11 - manual calibration mode

13 - sensor reactivation mode

3.10.1 Sensor reactivation

3.10.2 Standard measurement process

3.10.3 Grab sample process

Action

Screen

4-20 mA

Alarms

RS485

Address 0

Others

Frozen

Frozen

Frozen

Frozen

Adr 163=13

Action

Screen

4-20 mA

Alarms

RS485

Address 0

Others

Rinsing step

Frozen

Frozen

Frozen

Frozen

Measurement

Actual

Updated during

last minute of

measurement.

Updated during

last minute of

measurement.

Actual

Address 2, 4, 6, 8:

Updated at end of

the cycle

Action

Screen

4-20 mA

Alarms

RS485

Conc.

Other prog.

Address 0

Others

Phase 1

Frozen

Frozen

Event

Grab Sample

N/A

Frozen

Address 163=5

Phase 2
Phase 3
Phase 4
Phase 5

Actual

Actual

Actual