Section 2 specifications, 1 technical specifications – Hach-Lange POLYMETRON 9245 User Manual User Manual

Page 15

Advertising
background image

13

Section 2

Specifications

2.1 Technical specifications

Specifications are subject to change without notice.

Table 1 Technical specifications

PERFORMANCE SPECIFICATIONS

Measuring range

0 to 10,000 ppb freely programmable
0 to 200 ppm with K-Kit option

Accuracy

Non-cationic application:

± 0.1 ppb or ± 5% of reading, whichever is greater

Cationic application:

± 2 ppb or 5% reading, whichever is greater

Repeatability

< 0.02 ppb or 1.5% reading, whichever is greater within 10°C
variation

Response time

0.1 to 10 ppb T(90%) = 180 secs

Electrode type

pH glass electrode

Number of channels

1 channel

Interference phosphate 10 ppm

Measurement variation less 0.1 ppb.

Sample temperature interference

< 0.5% / °C

ENVIRONMENTAL REQUIREMENTS

Typical environment

Power station / indoor / demineralized water plant or
instrumentation room

Suspended solids

< 2 NTU, no oil, no grease.

Temperature range for storage

-20 to 60°C (2 to 140°F)

Relative humidity

10 to 80%

Ambient temperature

5 to 50°C (41 to 122°F)

Sample temperature variation

Stabilization in 10 mins from 15°C to 30°C

pH range of sample

Non-cationic application: 6 to 10 pH
Cationic application: 2 to 10 pH

Flow rate

6 to 9 L/hour

Pressure

0.2 to 6 bar (3 - 87 psi)

Acidity

Less than 250 ppm (equivalent CaCO

3

)

Power supply voltage fluctuation

± 10%

Over voltage category

2 (according to standard EN 61010-1)

Pollution degree

2 (according to standard CEI 664)

Altitude

< 2000 m

Measurement category

Cat II, Class 1 (overvoltage < 1500V)