4 reactivation frequency, User setup 7.2.4 reactivation frequency – Hach-Lange POLYMETRON 9245 User Manual User Manual

Page 76

Advertising
background image

74

User Setup

7.2.4 Reactivation frequency

Select the ACTIVAT. FREQ option to set the time period
between electrode reactivation. When this time period
expires, the electrode is reactivated automatically with an
injection of a small amount of reactivation solution.

Set the reactivation mode to either a Frequency or a fixed
date

.

If set to frequency, the recommended time period is 24 hours
so that the electrode is reactivated on a daily basis.

Note:

If this value is set to zero, then no electrode reactivation will

take place during the calibration process and as such the calibration
may be inaccurate. It is highly recommended to set this parameter to
24.

If set to a fixed date, define the day and time of the week
when reactivation takes place. Set the day of the week to an
asterisk if reactivation is not to take place on that day.

When the days have been defined, set the time in HH:MM:SS.

In the example screen shown left, reactivation will occur
automatically every Monday, Wednesday and Friday at 3
o’clock in the afternoon.

Advertising