Hach-Lange DR 6000 Basic User Manual User Manual

Page 160

Advertising
background image

English 160

Table 3 Options for Single Wavelength

Options

Description

More

For further Options

Recall
measurement data
symbol

Recall saved measurement data, wavelength scans or time
courses, refer to

Save, recall, send and delete data,

page 154

.

% Trans/Abs

Switches from % transmittance to concentration or
absorbance in reading mode.

λ

Wavelength

To input the wavelength setting. Use the alphanumeric
keypad to enter the reading wavelengths. A wavelength in
the range of 190–1100 nm can be entered.

Timer icon

This functions as a stopwatch. It helps to make sure that
the steps of an analysis are correctly timed (e. g. reaction
times, wait times, etc., can be exactly specified). When the
specified time has elapsed, a sound is emitted. The use of
the timer has no influence on the reading program.

Concentration
Factor: Off/On

Multiplication factor for converting absorbance values into
concentration values.

Concentration
Resolution

To select the number of decimal places.

Save as User
Program

To store the selected parameters as a User Program, refer
to

Single Wavelength (absorbance, concentration and

transmittance readings), page 159

.

Reading Mode

Single Reading: One measurement result is shown after
tapping on Read.
Continuous Readings: After the zero measurement, all
readings are displayed automatically and continuously.
Carousel 1 inch square: Optional measurement of
carousel insert with up to 5 square cells.
Carousel 1 cm square: Optional measurement of carousel
insert with up to 7 square cells.

Recall
measurement data

Recall saved measurement data, wavelength scans or time
courses, refer to

Save, recall, send and delete data,

page 154

.

Instrument Setup
mode

Basic data of the instrument, refer to

Save, recall, send and

delete data, page 154

.

Table 3 Options for Single Wavelength (Continued)

Options

Description