English 165

The Wavelength Scan is complete, if

the graph is shown full-size,

the scaling of the x-axis occurs automatically,

the Cursor functions in the vertical navigation bar are

highlighted.

a sound is emitted

To save the scan, press Options > Store symbol.

Time course of absorbance/transmittance

The Time Course mode is used to collect data in either absorbance or
transmittance for a user-specified length of time. This data can be
displayed in either a graph or a table.

Time course setup parameters

1.

Press Time Course mode in the Main Menu.

2.

Press Options to configure parameters.

Table 6 Time course options

Options

Description

More

For further options

Folder icon

To store the scan data

Time & Interval

To input the total time for data collection and the time interval
between the collection of data points

λ

To input the wavelength setting

View Table

To display readings in absorbance, transmittance or
concentration. This can be changed after sample data is
collected

Timer icon

This functions as a stopwatch. It helps to make sure that the
steps of an analysis are correctly timed (e. g. reaction times,
wait times, etc., can be exactly specified). When the
specified time has elapsed, a sound is emitted. The use of
the timer has no influence on the reading program.

Advertising