Hach-Lange DR 6000 Basic User Manual User Manual

Page 168

Advertising
background image

English 168

Entry invalid!

Password incorrect

Forgotten your
password?
Contact technical
support.

No valid data for these
parameters!

Data analysis not
possible, no
measurement data

Change the selection.

No valid data found!

View Data not possible in
data log

Change the selection.

No help function
present.

No measurement data
present!

Data analysis settings
cannot be configured
without measurement
data.

Change the selection.

Control range not
reached!

Data analysis limits not
reached

This is a warning notice.
The control limit set was
not reached.

Control range
exceeded!

Data analysis limits
exceeded.

This is a warning notice.
The control limit was
exceeded.

Concentration too high!

Calculated concentration
is higher than 999999

Dilute sample and
measure again

Over measuring range

The measured
absorbance is above the
calibration range of the
test

Dilute sample and
measure again

Under measuring range

The measured
absorbance is below the
calibration range of the
test

If possible, select a test
with a lower reading
range or use a cuvette
with a longer path length

Possible interference
by:

Interference Check

The analysis is possibly
erroneous due to
interferences.

Error displayed

Cause

Resolution

Possible interference
from:

Interference Check

The analysis is possibly
erroneous due to
interferences.

Next service is due!

Contact technical support
for an inspection of the
instrument.

Negative result!

The calculated result is
negative

Check concentration of
sample

Unstable lighting
conditions!

Avoid direct sunlight at
the measuring location.

System check
incorrect!

Measurement of air
values failed

Switch the instrument off
and then back on again.
If the system check is not
successful, contact
technical support.

Temperature too high.
Measurement not
possible!

Switch of the instrument
and allow it to cool for a
few minutes. If
necessary, move it to a
cooler place.

Update errors

An error occurred
when uploading the
instrument data.

Start the procedure again
or contact technical
support.

An error occurred when
reading from the USB
memory stick.

Start the procedure again
or contact technical
support.

An error occurred when
writing to the USB
memory stick.

Start the procedure again
or contact technical
support.

Please check on the
current update file.

Error during update.

Check the USB memory
stick.

Please contact
Customer Service.

Error during update.

Error displayed

Cause

Resolution