Welcome to the LTP-Director

13

Table: Connections for data transfer and amplifier control between USB-ME256 device,
MEA1060-BC amplifiers and data acquisition computer. Please note that each instance
of the LTP Director must be assigned to a specific data input.

Device / Connector Type

Cable

Device / Connector Type

MEA1060-BC A / Analog OUT

68-pin MCS high grade

USB-ME256 / Analog IN A

MEA1060-BC B / Analog OUT

68-pin MCS high grade

USB-ME256 / Analog IN B

MEA1060-BC C / Analog OUT

68-pin MCS high grade

USB-ME256 / Analog IN C

MEA1060-BC D / Analog OUT

68-pin MCS high grade

USB-ME256 / Analog IN D

MEA1060-BC A / Serial

Serial

Serial to USB Converter / COM 1 +
First Instance of LTP-Director

MEA1060-BC B / Serial

Serial

Serial to USB Converter / COM 2 +
Second Instance of LTP-Director

MEA1060-BC C / Serial

Serial

Serial to USB Converter / COM 3 +
Third Instance of LTP-Director

MEA1060-BC D / Serial

Serial

Serial to USB Converter / COM 4 +
Forth Instance of LTP-Director

Computer / USB port

USB

Serial to USB Converter / USB port

Computer / USB port

USB

USB-ME256 / USB port

1.

Connect the Analog IN A socket (68-pin MCS standard connector) of the USB-ME256 device

to the Analog OUT socket (68-pin MCS standard connector) of the first MEA1060-BC amplifier
via filter amplifier FA (not listed). This connection defines the MEA1060-BC amplifier as MEA A.
The connection from Analog IN B of the USB-ME256 device to the Analog OUT of the next
MEA1060-BC amplifier defines the amplifier as MEA B, and so on.

2.

Connect the Serial sockets of the MEA1060-BC amplifiers (serial cables) via Serial to USB converter

to the COM ports of the data acquisition computer. Please connect the MEA A amplifier to COM 1
of the computer, MEA B to COM 2, MEA C to COM 3, and MEA D to COM 4.

3.

The COM ports of the computer define the instances of the LTP-Director:

Important: Please make sure to connect MEA1060-BC amplifier A to COM 1 of the computer,
and start the first instance of LTP-Director. Connect MEA1060-BC amplifier B to COM 2 of the
computer, and start the second instance of LTP-Director. Connect MEA1060-BC amplifier C to
COM 3 of the computer, and start the third instance of LTP-Director. Connect MEA1060-BC
amplifier C to COM 4 of the computer, and start the fourth instance of LTP-Director.

Advertising