Gentec-EO PC-Gentec-EO (for MAESTRO and INTEGRA) User Manual

Page 50

Advertising
background image

PC-Gentec-

EO User’s Manual Revision 5.0

49

For example, a 12 milliwatts reading would be displayed like this:

Example

Response from Monitor

Command:

*CVU

0.012 <CR> <LF>

Continuous transmission of data

This command is used to send data to the serial port according to the data sampling setting. The
maximum transfer speed is 200Hz.

Text Command

Parameters

Return

CAU

None

Data in ASCII

Stop the CAU Command

This command is used to stop the real time transfer enabled by the CAU Command.

Text Command

Parameters

Return

CSU

None

Query New Value Ready

This command is used to check whether a new value is available from the device. Though
optional, its use is recommended when used with single pulse operations.

Text Command

Parameters

Return

NVU

Available/ Not Available <CR> <LF>

1/0

Example

Response from Monitor

Text Command:

*NVU <enter>

New Data Not Available <CR> <LF>

Note that the Query Current Value and Query Statistic Data commands will return the current
values from the device even if they have not been updated since the last query.

Get the Laser frequency

This command is used to get the laser frequency.