Text Command

Parameters

Return

GRR

None

Data in ASCII

Set the joulemeter binary mode

This command is used to set the monitor in binary or ASCII mode. Refer to section 3.4.2 for the
MAESTRO binary mode description.

Command

Parameters

Return

SS1

0= ASCII

1= Binary

Get Binary Joulemeter Mode

This command returns whether or not the binary joulemeter mode is activated for serial
communication. Refer to section 3.4.2 for the MAESTRO binary mode description.

Text Command

Parameters

Return

GBM

None

1: On
0: Off

Example

Response from Monitor

Command:

*GBM

Binary Joulemeter Mode : 0 \r\n

Set Analog Output

This command is used to enable or disable the output of the current value on the analog port of
the device.

Text Command

Parameters

Return

ANO

1 to Enable, 0 to Disable

MEASUREMENT COMMANDS -- SETUP

Set Personal Wavelength Correction

This command is used to specify the wavelength in nm being used on the detector. The
EEPROM in the detector contains measured spectral data for a wide range of wavelengths. If the
wavelength input by the user is different from the predefined list of wavelengths on the device, a
custom value is interpolated. Specifying zero as a wavelength or providing an out-of-bound value
as a parameter restores the default settings. A valid value is set between the lowest and highest
wavelengths supported by the device, and it should not be a floating point value. The input
parameter must have 5 digits. If the desired wavelength does not have 5 digits you must enter a
zero-padded number. For example 514nm, you must enter 00514.