Gentec-EO PC-Gentec-EO (for MAESTRO and INTEGRA) User Manual

Page 9

Advertising
background image

PC-Gentec-

EO User’s Manual Revision 5.0

8

Default Trigger Level

2 %

SOFTWARE
TRIGGER LEVEL

0.1% to 99.9%, 0.1% resolution

Transfer Rate for
MAESTRO

No missing points throughput serial frequency (power and energy):

500 Hz

For higher repetition rate, please use directly the MAESTRO

Refer to MAESTRO specification

4

:

Transfer Rate for
Integra

No missing points throughput serial frequency (power and energy):

5200 Hz

Frequency
Measurement

MAESTRO: Accurate frequency measurement up to 14kHz

0-1000 Hz: 0.1% accuracy

1000-2000 Hz: 0.5%

2000-14000 Hz: 1%

INTEGRA: Accurate frequency measurement up to 5.2kHz

0-5200 Hz: 1%

Statistics

Current value, Max, Min, Average, Std Dev., RMS stability, PTP stability,

Time, Pulse #.

General Specifications

Serial
Communication

Maestro and Integra USB

MAESTRO: RS-232

NOT AVAILABLE : Ethernet

Data Displays

Real Time, Scope, Averaging, Statistics, Digital Tuning Needle,

Histogram

User input correction
factors

1 multiplier and 1 offset (7 digits floating point)

Internet Upgrades

Downloadable link

Size

4.39 MB

Average RAM
allocation

9 to 23 MB

PC Recommended
Requirement

Dual Core Processor

3 Gb of RAM

Operating System
Compatibility

Windows 7

Windows Vista

Windows XP (not officially supported)

4

In a full size statistic’s windows, no scope display and no serial communication.

Advertising