PC-Gentec-

EO User’s Manual Revision 5.0

53

The following example sets multiplier = 33

Example

Response from Monitor

Text Command:

*MUL3300000

Or

*MUL3.3000e3

Get User Multiplier

This command returns the MAESTRO’s correction multiplier value.

Text Command

Parameters

Return

GUM

None

Current multiplier value

Example

Response from Monitor

Command:

*GUM

User Multiplier : 1 \r\n

Set User Offset

This command is used to set the value of the offset.

Text Command

Parameters

Return

OFF

8 characters numeral value

Default: 0

The following example sets offset to 1.5 milli.

Example

Response from Monitor

Text Command:

*OFF0.001500

OR

*OFF1.500e-3

The other option available is the Zero-offset. The Zero-offset operation is done first, before those
of the Multipliers and Offsets

Get User Offset

This command returns the MAESTRO’s correction offset value.

Text Command

Parameters

Return

GUO

None

Current offset value