Detector serial number (You must convert the

:0

002B

36 39

9 6

hexadecimal values in ASCII characters)

:0

002C

32 37

7 2

:0

002D

0 0

0000 = Null termination character

:0

002E

D70A

0.0200

Trigger Level LSB (between 0.001 and 0.999)

:0

002F

3CA3

Trigger Level MSB (between 0.001 and 0.999)

:0

0030

0001

1

Is autoscale mode on? LSB

:0

0031

0000

0

Is autoscale mode on? MSB

:0

0032

0000

0

Is anticipation on? LSB

:0

0033

0000

0

Is anticipation on? MSB

:0

0034

0000

0

Is zero offset on? LSB

:0

0035

0000

0

Is zero offset on? MSB

:0

0036

0000

1.0000

Correction Multiplier LSB

:0

0037

3F80

Correction Multiplier MSB

:0

0038

0000

0.0000

Correction Offset LSB

:0

0039

0000

Correction Offset MSB

:1

0000

0000

0

End of structure

Noise suppression (Integra only)

Sets or Queries the sampling size of the noise suppression. For joulemeter and UM only.

Text Command

Parameters

Return

AVG###

###: average size

none

Example

Response from Monitor

Send avg016 sets sampling size to 16 pulses, Integra replies
Ok.

Valid averages are powers of 2 from 1 to 128. If a non valid
argument is sent, An Error is returned and the value is left
unchanged.

Get current value with frequency (Integra only)

The *CTU can be called send current value with frequency.