0 physical inspection, 0 physical inspection -3 – ZOLL AED Pro Rev A User Manual

Page 15

Advertising
background image

Physical Inspection

9650-0309-01 Rev. A

ZOLL AED Pro Service Manual

1–3

1.0 Physical Inspection

Use this procedure to ensure that the unit shows no signs of damage or excessive wear.

Tools Needed

None.

Test Setup

None.

Test Procedure

Observe the following:

Pass

Fail

1.1

Is the unit clean?

‰

‰

1.2

Are all signs of wear reasonable and not excessive?

‰

‰

1.3

Are the front panel and housing undamaged and free of cracks?

‰

‰

1.4

Are input connectors clean and undamaged?

‰

‰

1.5

Are electrodes preconnected, is the cable free from damage, are the pads
sealed within the packaging, and are the electrodes within the expiration
date shown on the package?

‰

‰

1.6

Is the battery older than 18 months? When was the last time the battery
was tested?

‰

‰

1.7

If you are using an AED Pro 3-lead ECG cable, is it free from cuts or other
visible damage?

‰

‰