AED Pro Error Log Messages

The AED Pro unit records internal errors in an error log, which you can retrieve by using a
Windows-based personal computer with ZOLL Administration Software (ZAS).

To retrieve the error log, follow these steps:

After saving the error log file on the computer, you can use ZAS to clear the error log on the
AED Pro unit.

Table 2-4 lists critical error log messages.

Important: If any of these messages appears in the error log, contact ZOLL Technical Service

as described on page vii.

Step

Action

1

Turn off the AED Pro unit.

2

Start the ZOLL Administration Software (ZAS) on the computer.

3

Position the AED Pro unit so that its IrDA port has a clear line-of-sight with the computer’s
IrDA port and the ports are about 10 to 18 inches (25 to 45 cm) apart.

4

Press and hold the unit’s On/Off button for at least 5 seconds to start the unit in nonrescue
mode.

5

After the unit issues the message COMMUNICATIONS ESTABLISHED, refer to the ZAS
help for instructions on saving the unit’s history file and viewing the included error log.

Table 2-4. AED Pro Error Log Messages

Error

No.

Error Message

1

The patient relay is not working properly

2

Attempting to charge the capacitor while the SAFELINE in the CPLD is set - should not
charge

3

The capacitor takes too long to charge

4

The H-bridge test failed while the bridge was disabled. There should have been 0 current
measured.

5

VCAP1 and VCAP2 are not within a defined range of each other

6

The H-bridge test failed while the bridge was enabled - an invalid current reading was
taken during the positive test on the H-bridge

7

The H-bridge test failed while the bridge was enabled - an invalid current reading was
taken during the negative test on the H-bridge

8

Defib did not charge during the power-on charge test

9

The Defib task is not receiving A2D vCap1 event samples - needed for charging

10

FnSafety has determined that the charge has dropped below a specified value during the
holding of charge.

11

The capacitor failed to charge

12

The capacitor over-charged

13

The capacitor contained an unsafe voltage in an idle state