Troubleshooting the aed pro unit, Troubleshooting the aed pro unit -2 – ZOLL AED Pro Rev A User Manual

Page 24

Advertising
background image

C

HAPTER

2 T

ROUBLESHOOTING

2–2

www.zoll.com

9650-0309-01 Rev. A

Troubleshooting the AED Pro Unit

Table 2-1 lists general issues with the unit and their associated corrective action.

First, try the recommendations given under “Operator Action.” If these steps do not remedy the
problem, follow the suggestions under “Technical Action.”

For information on contacting ZOLL Technical Service, refer to page vii.

Table 2-1. General Issues

Symptom

Operator Action

Technical Action

Unit beeps or displays a red
“X” while turned off.

— Turn the unit on. Follow the
prompts to resolve the problem.

If the beeping continues,
take the unit out of service.

Contact ZOLL Technical
Service.

Ready indicator shows a red
“X” while the device is
powered on.

— Turn the unit off and then on again.
Follow the prompts to resolve the
problem.

If the red “X” remains, take
the unit out of service.

Contact ZOLL Technical
Service.

Power-on self-test failed.

— Follow the prompts to resolve the
problem.

If the unit continues to fail,
take it out of service.

Contact ZOLL Technical
Service.

Unexpected shutdown in
clinical mode.

Note: In clinical mode, the unit

automatically powers off if it
does not detect a patient
connection within 10 minutes
(configurable).

— Turn the unit on. Follow the
prompts to resolve the problem.

If the unit continues to fail,
take it out of service.

Contact ZOLL Technical
Service.

Unexpected shutdown in
nonrescue mode.

Note: The unit automatically powers

off if an established IrDA
connection is lost.

— Press and hold the On/Off button
for at least 5 seconds. Follow the
prompts to resolve the problem.

If the unit still is not ready for
use, take it out of service.

Contact ZOLL Technical
Service.

;

CHANGE BATTERY

— Replace the battery pack with a
fully charged battery pack as soon as
possible.

None.

;

PLUG IN CABLE

— Ensure that the electrode cable is
properly connected to the unit.

— Remove the cable and check for
bent or broken pins.

— Replace the electrode cable.

None.

;

ANALYSIS HALTED
KEEP PATIENT STILL

— Keep the patient still during ECG
analysis. If transporting the patient by
stretcher or vehicle, stop all patient
movement during ECG analysis.

None.

Advertising