ZOLL AED Pro Rev A User Manual

Page 19

Advertising
background image

Preventive Maintenance Test

9650-0309-01 Rev. A

ZOLL AED Pro Service Manual

1–7

Turn off the AED Pro unit.
Set the analyzer to NSR (normal sinus rhythm).

2.5

Turn on the AED Pro unit.

;

UNIT OK

;

ADULT PADS

;

DON’T TOUCH PATIENT,
ANALYZING
.

;

NO TREATMENT ADVISED
(or NO SHOCK ADVISED)

These three prompts are
optional:
OPEN AIRWAY
CHECK BREATHING
GIVE TWO BREATHS

;

CHECK CIRCULATION
(or CHECK PULSE)

;

IF NO CIRCULATION
START CPR

(or IF NO PULSE START
CPR
)

‰

‰

‰

2.6

Lift the puck that is attached to
the universal adapter cable.
Move it up and down very
slightly (no more than 1 inch or
2.5 cm) and continue to do so.

On the display screen, prevent
the bar on the chest
compression gauge from
extending into the area between
the lower two lines.

Unit emits spaced tones as a

guide for performing chest
compressions at the proper
rate.

;

PUSH HARDER

‰

‰

‰

Test

Action

Expected Result(s)

Pass

Fail

N/A

Advertising