0 manual self-test, 0 manual self-test -10 – ZOLL AED Pro Rev A User Manual

Page 22

Advertising
background image

C

HAPTER

1 M

AINTENANCE

T

ESTS

1–10

www.zoll.com

9650-0309-01 Rev. A

5.0 Manual Self-Test

Use this test to initiate the unit’s self-diagnostic procedure, which verifies the following:

Battery usage since last reset.

Defibrillation electrodes are properly connected.

ECG signal acquisition and processing electronics are functional.

Defibrillator charge and discharge circuitry (operational at 2 joules) is functional.

Microprocessor hardware and software are functional.

CPR monitoring circuitry and compression depth sensor are functional (when ZOLL
CPR-D•padz are attached).

Audio output circuitry is functional.

Display screen is functional.

Tools Needed

None.

Test Setup

Turn off the AED Pro unit.

If they are not already in place, preconnect defibrillation electrodes to the unit.

Test Procedure

Test

Action

Expected Result(s)

Pass

Fail

5.1

Press and hold the On/Off button
for at least 5 seconds.

Within 10 seconds after the unit is

turned on, the Ready indicator
displays a green check.

;

UNIT OK

‰

‰

If an IrDA connection is established, turn off the unit; otherwise, the unit shuts down
automatically.