0 manual self-test, 0 manual self-test -10 – ZOLL AED Pro Rev A User Manual

Page 22

Advertising
background image

C

HAPTER

1 M

AINTENANCE

T

ESTS

1–10

www.zoll.com

9650-0309-01 Rev. A

5.0 Manual Self-Test

Use this test to initiate the unit’s self-diagnostic procedure, which verifies the following:

•

Battery usage since last reset.

•

Defibrillation electrodes are properly connected.

•

ECG signal acquisition and processing electronics are functional.

•

Defibrillator charge and discharge circuitry (operational at 2 joules) is functional.

•

Microprocessor hardware and software are functional.

•

CPR monitoring circuitry and compression depth sensor are functional (when ZOLL
CPR-D•padz are attached).

•

Audio output circuitry is functional.

•

Display screen is functional.

Tools Needed

•

None.

Test Setup

•

Turn off the AED Pro unit.

•

If they are not already in place, preconnect defibrillation electrodes to the unit.

Test Procedure

Test

Action

Expected Result(s)

Pass

Fail

5.1

Press and hold the On/Off button
for at least 5 seconds.

• Within 10 seconds after the unit is

turned on, the Ready indicator
displays a green check.

;

UNIT OK

‰

‰

If an IrDA connection is established, turn off the unit; otherwise, the unit shuts down
automatically.