ZOLL AED Pro Rev A User Manual
Page 17
Preventive Maintenance Test
9650-0309-01 Rev. A
ZOLL AED Pro Service Manual
1–5
Test Procedure
Test
Action
Expected Result(s)
Pass
Fail
N/A
2.1
Turn on the AED Pro unit.
Note: Some AED Pro units are
configured to start with CPR;
therefore, these prompts will not
be heard.
;
UNIT OK
;
ADULT PADS
;
CHECK RESPONSIVENESS
(or CHECK PATIENT)
;
DON’T TOUCH PATIENT,
ANALYZING
;
TREATMENT ADVISED
(or SHOCK ADVISED)
;
DON’T TOUCH PATIENT
;
PRESS FLASHING
TREATMENT BUTTON
(or PRESS FLASHING
SHOCK BUTTON)
2.2
Press the Shock button
(first shock).
• Analyzer indicates a
defibrillator discharge at the
configured first-shock energy
level (120, 150, or 200
joules) ±15%.
;
TREATMENT DELIVERED
(or SHOCK DELIVERED)
• An updated shock count
appears on the display.
;
DON’T TOUCH PATIENT,
ANALYZING
;
TREATMENT ADVISED
(or SHOCK ADVISED)
;
DON’T TOUCH PATIENT
;
PRESS FLASHING
TREATMENT BUTTON
(or PRESS FLASHING
SHOCK BUTTON)