ZOLL AED Pro Rev A User Manual

Page 17

Advertising
background image

Preventive Maintenance Test

9650-0309-01 Rev. A

ZOLL AED Pro Service Manual

1–5

Test Procedure

Test

Action

Expected Result(s)

Pass

Fail

N/A

2.1

Turn on the AED Pro unit.

Note: Some AED Pro units are
configured to start with CPR;
therefore, these prompts will not
be heard.

;

UNIT OK

;

ADULT PADS

;

CHECK RESPONSIVENESS
(or CHECK PATIENT)

;

DON’T TOUCH PATIENT,
ANALYZING

;

TREATMENT ADVISED
(or SHOCK ADVISED)

;

DON’T TOUCH PATIENT

;

PRESS FLASHING
TREATMENT BUTTON

(or PRESS FLASHING
SHOCK BUTTON
)

‰

‰

‰

2.2

Press the Shock button
(first shock).

Analyzer indicates a

defibrillator discharge at the
configured first-shock energy
level (120, 150, or 200
joules) ±15%.

;

TREATMENT DELIVERED
(or SHOCK DELIVERED)

An updated shock count

appears on the display.

;

DON’T TOUCH PATIENT,
ANALYZING

;

TREATMENT ADVISED
(or SHOCK ADVISED)

;

DON’T TOUCH PATIENT

;

PRESS FLASHING
TREATMENT BUTTON

(or PRESS FLASHING
SHOCK BUTTON
)

‰

‰

‰

Advertising