ZOLL AED Pro Rev A User Manual

Page 18

Advertising
background image

C

HAPTER

1 M

AINTENANCE

T

ESTS

1–6

www.zoll.com

9650-0309-01 Rev. A

2.3

Press the Shock button
(second shock).

Note: If your AED Pro unit is not
configured for stack shocks,
skip to test 2.5.

Analyzer indicates a

defibrillator discharge at the
configured second-shock
energy level (120, 150, or
200 joules) ±15%.

;

TREATMENT DELIVERED
(or SHOCK DELIVERED)

An updated shock count

appears on the display.

;

DON’T TOUCH PATIENT,
ANALYZING

;

TREATMENT ADVISED
(or SHOCK ADVISED)

;

DON’T TOUCH PATIENT

;

PRESS FLASHING
TREATMENT BUTTON

(or PRESS FLASHING
SHOCK BUTTON
)

‰

‰

‰

2.4

Press the Shock button
(third shock).

Note: If your AED Pro unit is not
configured for stack shocks,
skip to test 2.5.

Analyzer indicates a

defibrillator discharge at the
configured third-shock
energy level (120, 150, or
200 joules) ±15%.

;

TREATMENT DELIVERED
(or SHOCK DELIVERED)

An updated shock count

appears on the display.

These three prompts are
optional:
OPEN AIRWAY
CHECK BREATHING
GIVE TWO BREATHS

;

CHECK CIRCULATION
(or CHECK PULSE)

;

IF NO CIRCULATION
START CPR

(or IF NO PULSE START
CPR
)

‰

‰

‰

Test

Action

Expected Result(s)

Pass

Fail

N/A

Advertising