ZOLL AED Pro Rev A User Manual
Page 18
C
HAPTER
1 M
AINTENANCE
T
ESTS
1–6
www.zoll.com
9650-0309-01 Rev. A
2.3
Press the Shock button
(second shock).
Note: If your AED Pro unit is not
configured for stack shocks,
skip to test 2.5.
• Analyzer indicates a
defibrillator discharge at the
configured second-shock
energy level (120, 150, or
200 joules) ±15%.
;
TREATMENT DELIVERED
(or SHOCK DELIVERED)
• An updated shock count
appears on the display.
;
DON’T TOUCH PATIENT,
ANALYZING
;
TREATMENT ADVISED
(or SHOCK ADVISED)
;
DON’T TOUCH PATIENT
;
PRESS FLASHING
TREATMENT BUTTON
(or PRESS FLASHING
SHOCK BUTTON)
2.4
Press the Shock button
(third shock).
Note: If your AED Pro unit is not
configured for stack shocks,
skip to test 2.5.
• Analyzer indicates a
defibrillator discharge at the
configured third-shock
energy level (120, 150, or
200 joules) ±15%.
;
TREATMENT DELIVERED
(or SHOCK DELIVERED)
• An updated shock count
appears on the display.
These three prompts are
optional:
— OPEN AIRWAY
— CHECK BREATHING
— GIVE TWO BREATHS
;
CHECK CIRCULATION
(or CHECK PULSE)
;
IF NO CIRCULATION
START CPR
(or IF NO PULSE START
CPR)
Test
Action
Expected Result(s)
Pass
Fail
N/A