Performing a manual battery test – ZOLL SurePower Rev B Charger Station User Manual
Page 27
Advertising

Testing Batteries (Manual Procedure)
9650-0535-01 Rev. B
SurePower Charger Station
Operator’s Guide
3-5
Performing a Manual Battery Test
To perform a manual battery test, follow this procedure:
Step
Action
1.
Insert the battery into the charging bay and press the Test Button on the bay’s
control panel.
The charger lights the Test LED and begins testing the battery. The Test LED will
remain lit for the duration of the test.
When the battery is fully charged and has passed a factory-configured minimum
capacity test, the charger lights the Ready and Test LED’s.
If the charger determines that the battery has insufficient capacity, it lights the
Fault
and Test LED’s.
Advertising