Chapter 4 maintenance and troubleshooting, Inspection, Cleaning – ZOLL SurePower Rev B Charger Station User Manual

Page 33

Advertising
background image

Inspection

9650-0535-01 Rev. B

SurePower Charger Station

Operator’s Guide

4-1

Chapter 4

Maintenance and

Troubleshooting

This chapter describes a series of non-technical operational checks that you should perform
annually to ensure the proper operation of the

SurePower Charger Station

.

Inspection

Begin by inspecting the charger to ensure that it’s clean (with no fluid spills). Make sure that
nothing has been placed on top of the charger and that there are at least four inches of clear
space around the charger so that it can dissipate heat effectively.

You can then perform the following quick, non-technical operational checks:

1.

Inspect for cracks or damage to the charger housing.

2.

Inspect the AC cord for severe kinks or exposed wires.

3.

Connect the charger to live AC mains. The charger will beep, turn on the fan, and light all
LEDs for several seconds. The Power LED will remain lit after the other LEDs go off.

4.

With all battery bays empty, check that all LEDs on the control panel except the Power LED
are off.

5.

Press the Test button under each bay to verify that the bays are fully operational (for more
information, see the section, “Testing Battery Bays,” in Chapter 3).

6.

Insert a battery into each bay. Verify that the appropriate LEDs in each bay immediately
illuminate.

Cleaning

Clean the

SurePower Charger Station

with a soft cloth, mild soap, and water.

Do not immerse any part of the

SurePower Charger Station

in water. Do not use alcohol or

ketones (MEK, acetone, etc.). Do not autoclave the

SurePower Charger Station

.

Advertising