Electromagnetic immunity declaration (eid) – ZOLL SurePower Rev B Charger Station User Manual

Page 42

Advertising
background image

M

ANUFACTURER

S

D

ECLARATION

-- E

LECTROMAGNETIC

E

MISSIONS

AND

I

MMUNITY

-G

UIDANCE

AND

A-2

www.zoll.com

9650-0535-01 Rev. B

Electromagnetic Immunity Declaration (EID)

Immunity Test

IEC 60601 Test
Level

Compliance Level

Electromagnetic
Environment - Guidance

Electrostatic
discharge (ESD)
IEC 61000 4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.

Electrical fast
transient/burst
IEC 61000 4-4

± 2 kV for power
supply lines

± 1 kV for input/
output lines

± 2 kV for power
supply lines

± 1 kV

Mains power quality
should be that of a typical
commercial or hospital
environment.

Surge
IEC 61000 4-5

± 1 kV differential
mode

± 2 kV for common
mode

± 1 kV differential
mode

± 2 kV for common
mode

Mains power quality
should be that of a typical
commercial or hospital
environment.

Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11

<5% U

T

(>95% dip

in U

T

) for 0.5 cycle

40% U

T

(60% dip in

U

T

) for 5 cycles

70% U

T

(30% dip in

U

T

) for 25 cycles

<5% U

T

(>95% dip

in U

T

) for 5 sec.

<5% U

T

(>95% dip

in U

T

) for 0.5 cycle

40% U

T

(60% dip in

U

T

) for 5 cycles

70% U

T

(30% dip in

U

T

) for 25 cycles

<5% U

T

(>95% dip

in U

T

) for 5 sec.

Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the

SurePower Charger

Station

unit requires

continued operation
during power mains
interruptions, it is
recommended that the

SurePower Charger
Station

unit be powered by

an uninterruptable power
supply or a battery.

Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8

3 A/m

3 A/m

Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment.

U

T

is the AC mains voltage prior to application of the test level.