Electromagnetic immunity declaration (eid) – ZOLL SurePower Rev B Charger Station User Manual
Page 42
M
ANUFACTURER
’
S
D
ECLARATION
-- E
LECTROMAGNETIC
E
MISSIONS
AND
I
MMUNITY
-G
UIDANCE
AND
A-2
www.zoll.com
9650-0535-01 Rev. B
Electromagnetic Immunity Declaration (EID)
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic
Environment - Guidance
Electrostatic
discharge (ESD)
IEC 61000 4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000 4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2 kV for power
supply lines
± 1 kV
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000 4-5
± 1 kV differential
mode
± 2 kV for common
mode
± 1 kV differential
mode
± 2 kV for common
mode
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip
in U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip
in U
T
) for 5 sec.
<5% U
T
(>95% dip
in U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip
in U
T
) for 5 sec.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the
SurePower Charger
Station
unit requires
continued operation
during power mains
interruptions, it is
recommended that the
SurePower Charger
Station
unit be powered by
an uninterruptable power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment.
U
T
is the AC mains voltage prior to application of the test level.