10 emc conformity standards, 11 ped (pressure equipment directive) – Yokogawa EJA110A User Manual

Page 19

Advertising
background image

IM 01C21B01-01E

2-11

2. HANDLING CAUTIONS

2.10 EMC Conformity Standards

EN61326-1 Class A, Table 2 (For use in industrial
locations)

EN61326-2-3

CAUTION

This instrument is a Class A product, and it is
designed for use in the industrial environment.
Please use this instrument in the industrial
environment only.

NOTE

YOKOGAWA recommends customer to apply
the Metal Conduit Wiring or to use the twisted
pair Shield Cable for signal wiring to conform the
requirement of EMC Regulation, when customer
installs the EJA Series Transmitters to the plant.

2.11 PED (Pressure Equipment

Directive)

(1) General

• EJA series of pressure transmitters are categorized

as pressure accessories under the vessel section of
this directive 97/23/EC, which corresponds to
Article 3, Paragraph 3 of PED, denoted as Sound
Engineering Practice (SEP).

• EJA130A, EJA440A, EJA510A, and EJA530A can

be used above 200 bar and therefore considered as
a part of a pressure retaining vessel where category
lll, Module H applies. These models with option
code /PE3 conform to that category.

(2) Technical Data

• Models without /PE3
Article 3, Paragraph 3 of PED, denoted as Sound

Engineering Practice (SEP) .

• Models with /PE3
Module: H
Type of Equipment: Pressure Accessory - Vessel
Type of Fluid: Liquid and Gas
Group of Fluid: 1 and 2

Model

EJA110A

EJA120A

EJA130A

EJA130A

With code /PE3

EJA310A

EJA430A

EJA440A

EJA440A

With code /PE3

EJA510A

EJA510A

With code /PE3

EJA530A

EJA530A

With code /PE3

Category

*2

Article 3, paragraph 3

(SEP)

Article 3, paragraph 3

(SEP)

Article 3, paragraph 3

(SEP)

III

Article 3, paragraph 3

(SEP)

Article 3, paragraph 3

(SEP)

Article 3, paragraph 3

(SEP)

III

Article 3, paragraph 3

(SEP)

III

Article 3, paragraph 3

(SEP)

III

PS-V(bar-L)

1.6

0.005

4.2

4.2

1.6

1.6

50

50

50

50

50

50

V(L)

0.01

0.01

0.01

0.01

0.01

0.01

0.01

0.01

0.01

0.01

0.01

0.01

PS(bar)

*1

160

0.5

420

420

160

160

500

500

500

500

500

500

*1: PS is maximum allowable pressure for vessel itself.
*2: Referred to Table 1 covered by ANNEX II of EC Directive
on Pressure Equipment Directive 97/23/EC

T0299.EPS

(3) Operation

CAUTION

• The temperature and pressure of fluid should

be applied under the normal operating condi-
tion.

• The ambient temperature should be applied

under the normal operating condition.

• Please pay attention to prevent the excessive

pressure like water hammer, etc. When water
hammer is to be occurred, please take mea-
sures to prevent the pressure from exceeding
PS by setting the safety valve, etc. at the
system and the like.

• When external fire is to be occurred, please

take safety measures at the device or system
not to influence the transmitters.

Advertising