2 zero point adjustment, Zero point adjustment -2 – Yokogawa EJA110A User Manual

Page 34

Advertising
background image

IM 01C21B01-01E

7-2

7. OPERATION

᭿

Confirming that Transmitter is Operating
Properly

Using the BT200

• If the wiring system is faulty, ‘communication error’

appears on the display.

• If the transmitter is faulty, ‘SELF CHECK ERROR’

appears on the display.

communication error

PARAM
C60:SELF CHECK
ERROR

Communication error
(Faulty wiring)

Self-diagnostic error
(Faulty transmitter)

DATA

DIAG

PRNT

ESC

F0702.EPS

Using the integral indicator

• If the wiring system is faulty, the display stays blank.
• If the transmitter is faulty, an error code will appear

on the display according to the nature of the error.

Self-diagnostic error on
the integral indicator
(Faulty transmitter)

F0703.EPS

NOTE

If any of the error indications above appears on
the display of the integral indicator or BT200,
refer to Subsection 8.5.2 for corrective action.

᭿

Verify and Change Transmitter Parameter
Setting and Values

The following parameters are the minimum settings
required for operation. The transmitter has been
shipped with these parameters. To confirm or change
the values, see Subsection 8.3.3.

• Measuring range ....... See Subsection 8.3.3 (2)
• Output/integral indicator mode .........

See Subsection 8.3.3 (4)

• Operation mode ......... See Subsection 8.3.3 (9)

7.2 Zero Point Adjustment

Adjust the zero point after operating preparation is
completed.

IMPORTANT

Do not turn off the power to the transmitter
immediately after a zero adjustment. Powering
off within 30 seconds after a zero adjustment will
return the adjustment back to the previous
settings.

The zero point adjustment can be made in either way:
using the zero-adjustment screw of the transmitter or
the BT200 operation.

For output signal checking, display the parameter

A10:

OUTPUT (%)

in the BT200.

Output signal (%)
display

PARAM
A10:OUTPUT(%)
0.0 %
A11:ENGR OUTPUT

A20:AMP TEMP

DATA

DIAG

PRNT

ESC

F0704.EPS

Zero-adjustment
screw

BT200

Zero-adjustment Screw

᭿

Using the Transmitter Zero-adjustment
Screw

Before adjusting a screw, check that the parameter

J20: EXT ZERO ADJ

displays

ENABLE.

See

Subsection 8.3.3 (15) for the setting procedure.

Use a slotted screwdriver to turn the zero-adjustment
screw. Turn the screw clockwise to increase the output
or counterclockwise to decrease the output. The zero
point adjustment can be made with a resolution of
0.01% of the setting range. Since the degree of zero
adjustments varies with the screw turning speed, turn
the screw slowly for fine adjustment and quickly for
coarse adjustment.