10. GENERAL SPECIFICATIONS

10. GENERAL SPECIFICATIONS

10.1

Standard Specifications

Refer to IM 01C22T02-01E for F

OUNDATION

Fieldbus communication type and IM 01C22T03-
00E for PROFIBUS PA communication type
marked with “

”.

Performance Specifications

See General Specifications sheet, GS 01C21B01-
00E, GS 01C21B03-00E, and GS 01C21B04-00E.

Functional Specifications

Span & Range Limits

EJA110A and EJA130A:

L

M

Measurement

Span

and Range

Span

Range

Span

Range

inH

2

O

(/D1)

2 to 40

–40 to 40

4 to 400

–400 to

400

mbar
(/D3)

5 to 100

mmH

2

O

(/D4)

50 to 1000

–100 to 100

10 to 1000

–1000 to

1000

–1000 to

1000

100 to 10000

–10000 to

10000

kPa

0.5 to 10

–10 to 10

1 to 100

–100 to

100

H

Span

Range

20 to 2000

–2000 to

2000

50 to 5000

–5000 to

5000

0.05 to

5 kgf/cm

2

–5 to

5 kgf/cm

2

5 to 500

–500 to

500

T1001.EPS

V*

0.14 to

14 MPa

20 to

2000 psi

1.4 to

140 bar

1.4 to

140

kgf/cm

2

–0.5 to

14 MPa

–71 to

2000 psi

–5 to

140 bar

–5 to

140 kgf/cm

2

Span

Range

* For Wetted parts material code other than S, the ranges

are 0 to 14 MPa, 0 to 2000 psi, 0 to 140 bar, and 0 to
140 kgf/cm

2

.

EJA120A:

E

Measurement

Span

and Range

Span

Range

inH

2

O

(/D1)

0.4 to 4

–4 to 4

mbar
(/D3)

1 to 10

mmH

2

O

(/D4)

10 to 100

–10 to 10

–100 to 100

kPa

0.1 to 1

–1 to 1

T1002.EPS

URL is define as the Upper Range Limit from the table
above.

Zero Adjustment Limits:

Zero can be fully elevated or suppressed, within
the Lower and Upper Range Limits of the capsule.

External Zero Adjustment “

”:

External zero is continuously adjustable with
0.01% incremental resolution of span. Span may
be adjusted locally using the digital indicator with
range switch.

Output “

”:

Two wire 4 to 20 mA DC output with digital
communications, linear or square root program-
mable. BRAIN or HART FSK protocol are superim-
posed on the 4 to 20 mA signal.

Failure Alarm:

Output status at CPU failure and hardware error;

Up-scale: 110%, 21.6 mA DC or more(standard)
Down-scale: -5%, 3.2 mA DC or less

-2.5%, 3.6 mA DC or less(Optional

code /F1)

Note: Applicable for Output signal code D and E

Damping Time Constant (1st order):

The sum of the amplifier and capsule damping
time constant must be used for the overall time
constant. Amp damping time constant is adjustable
from 0.2 to 64 seconds.

Time Constant (approx. sec)

0.4

0.3

0.3

0.3

0.2

L

M

H

V

E

T1003.EPS

Capsule (Silicone Oil)

Ambient Temperature Limits:

* Safety approval codes may affect limits.

EJA110A, EJA130A

–40 to 85

°

C (–40 to 185

°

F)

–30 to 80

°

C (–22 to 176

°

F) with LCD Display

EJA120A

–25 to 80

°

C (–13 to 176

°

F)

Process Temperature Limits:

* Safety approval codes may affect limits.

EJA110A, EJA130A

–40 to 120

°

C (–40 to 248

°

F)

EJA120A

–25 to 80

°

C (–13 to 176

°

F)

Maximum Overpressure:

See General Specifications sheet.

Working Pressure Limits (Silicone Oil)

Maximum Pressure Limit:

See ‘Model and Suffix Codes’

Minimum Pressure Limit:

Capsule

Pressure

L, M, H (EJA110A)

See Figure 1.

M, H (EJA130A)

See Figure 1.

E (EJA120A)

–50 kPa {–7.25 psig}