4 replacing the process connector gaskets, 5 troubleshooting, 1 basic troubleshooting – Yokogawa EJA110A User Manual

Page 62: Replacing the process connector gaskets -6, Troubleshooting -6, Basic troubleshooting -6

Advertising
background image

IM 01C21B01-01E

9-6

9. MAINTENANCE

F0904.EPS

Transmitter section

Pressure-detector section

Cover flange

Capsule gasket

Flange bolt

Allen screw

Nut

Figure 9.4.3 Removing and Mounting the Pressure-

detector Section

9.4.4 Replacing the Process Connector

Gaskets

This subsection describes process connector gasket
replacement. (See Figure 9.4.4.)

(a) Loosen the two bolts, and remove the process

connectors.

(b) Replace the process connector gaskets.
(c) Remount the process connectors. Tighten the bolts

securely and uniformly to a torque of 39 to 49 N·m
{4 to 5 kgf·m}, and verify that there are no pressure
leaks.

F0905.EPS

Bolt

Process connector

Process connector gasket

Figure 9.4.4 Removing and Mounting the Process

Connector

9.5 Troubleshooting

If any abnormality appears in the measured values, use
the troubleshooting flow chart below to isolate and
remedy the problem. Since some problems have
complex causes, these flow charts may not identify all.
If you have difficulty isolating or correcting a problem,
contact Yokogawa service personnel.

9.5.1 Basic Troubleshooting

First determine whether the process variable is actually
abnormal or a problem exists in the measurement
system.

If the problem is in the measurement system, isolate
the problem and decide what corrective action to take.

This transmitter is equipped with a self-diagnostic
function which will be useful in troubleshooting; see
Section 8.5 for information on using this function.

Abnormalities appear in measurement.

: Areas where self-diagnostic offers support

Is process variable

itself abnormal?

Inspect the

process system.

Isolate problem in

measurement system.

Does problem exist in

receiving instrument?

Check/correct

environmental conditions.

Inspect receiver.

Check transmitter.

Check/correct operating

conditions.

Measurement system problem

F0906.EPS

YES

NO

NO

YES

Environmental conditions

Operating conditions

Transmitter itself

Figure 9.5.1 Basic Flow and Self-Diagnostics

Advertising