7. BRAIN TERMINAL BT200 OPERATION

(4) Output Mode and Integral Indicator Dis-

play Mode Setup (C40: OUTPUT MODE)

The mode setting for the output signal and the integral
indicator coordinate as shown in the table below.

BT200 Display

Output Mode

Integral Indicator

Display Mode

T0708.EPS

Linear

Linear

Square root

OUT: LIN DSP: LIN

OUT: LIN DSP: SQR

OUT: SQR DSP: SQR

Linear

Square root

Square root

This mode is set as specified in the order when the
instrument is shipped. Follow the procedure below to
change the mode.

If the instrument is equipped with an integral indicator
and the display mode is “square root”, “

” is dis-

played on the integral indicator.

For details, see Chapter 3.

• Example: Set output mode to Linear and

display mode to Square root.

ESC

SET
C40:OUTPUT MODE
OUT:LIN DSP:SQR

FEED

NO

ESC

SET
C40:OUTPUT MODE
OUT:LIN DSP:LIN
<OUT:LIN DSP:LIN >
<OUT:LIN DSP:SQR.>
<OUT:SQR DSP:SQR >

Use the or key

to select “OUT: LIN, DSP: SQR.”

Press the key twice to

enter the setting.

Press the (OK) key.

F0716.EPS

(5) Output Signal Low Cut Mode Setup

(D10: LOW CUT, D11: LOW CUT MODE)

Low cut mode can be used to stabilize the output
signal near the zero point. The low cut point can be set
in a range from 0 to 20% of output. (Hysteresis of cut
point:

±

1%)

Either “

LINEAR

” or “

ZERO

” can be selected as the

low cut mode.

(%)

50

20

0

50 (%)

Output

Low cut mode “LINEAR”

Input

(%)

50

20

0

50 (%)

Low cut mode “ZERO”

LOW CUT at 20%

Input

F0717.EPS

• Example: Change the low cut setting range

from 10% to 20%, and the low
cut mode from LINEAR to ZERO.

FEED

NO

OK

SET
D11:LOW CUT MODE
ZERO

Use the or key

to select “ZERO.”

Press the key twice to

enter the setting.

Press the (OK) key.

F0718.EPS

FEED

NO

OK

SET
D10:LOW CUT
20.0 %

ESC

SET
C11:LOW CUT MODE
LINEAR
< LINEAR >
< ZERO >

DATA

DIAG

PRNT

ESC

PARAM
D10:LOW CUT
20.0 %
D11:LOW CUT MODE
ZERO
D20:DISP SELECT
NORMAL %

CLR

ESC

SET
D10:LOW CUT
10.0 %
+ 20.0

Set “20.”

Press the key twice to

enter the setting.

Press the (OK) key.

Next, the [D11: LOW CUT MODE]

setting panel is displayed.