4 disassembly and reassembly, 1 replacing the integral indicator, 4 disassembly and reassem- bly – Yokogawa EJA118 User Manual

Page 50

Advertising
background image

IM 1C22H1-01E

8-3

8. MAINTENANCE

8.4 Disassembly and Reassem-

bly

This section describes procedures for disassembly and
reassembly for maintenance and component replace-
ment.

Always turn OFF power and shut off and release
pressures before disassembly. Use proper tools for all
operations. Table 8.4.1 shows the tools required.

Table 8.4.1 Instruments Required for Calibration

Tool

Remarks

Quantity

Phillips screwdriver
Slotted screwdriver

Allen wrenches

Wrench

Torque wrench

Adjustable wrench

Socket wrench

Socket driver

Tweezers

JIS B4633, No. 2

JIS B4648
One each, nominal 3 and
5 mm Allen wrenches
Width across flats, 17 mm

Width across flats, 16 mm
Width across flats, 5.5 mm

T0802.EPS

1
1
2

1
1
1
1
1
1

CAUTION

Precautions for CENELEC, SAA and JIS

Flameproof Type Transmitters

• Flameproof type transmitters must be, as a

rule, removed to a non-hazardous area for
maintenance and be disassembled and reas-
sembled to the original state.

• Two covers are locked by each of an Allen

head bolt (shrouding bolt) on the flameproof
type transmitters. When a shrouding bolt is
driven clockwise by an Allen wrench, it is going
in and cover lock is released, and then a cover
can be opened.
When a cover is closed it should be locked by
a shrouding bolt without fail. Tighten the
shrouding bolt to a torque of 0.7 N·m.

Shrouding bolt

Shrouding bolt

F0802.EPS

Figure 8.4.1 Shrouding Bolts

8.4.1 Replacing the Integral Indicator

This subsection describes the procedure for replacing
an integral indicator. (See Figure 8.4.2)

CAUTION

Precautions for JIS Flameproof Type Trans-

mitters

Users are prohibited by law from modifying the
construction of a flameproof type transmitter.
Thus the user is prohibited from using a flame-
proof type transmitter with its integral indicator
removed, or from adding an integral indicator to
a transmitter. If such modification is absolutely
required, contact Yokogawa.

᭿

Removing the Integral Indicator

1) Remove the cover.
2) Supporting the integral indicator by hand, loosen

its two mounting screws.

3) Dismount the LCD board assembly from the CPU

assembly.
When doing this, carefully pull the LCD board
assembly straight forward so as not to damage the
connector between it and the CPU assembly.

᭿

Attaching the Integral Indicator

1) Align both the LCD board assembly and CPU

assembly connectors and engage them.

2) Insert and tighten the two mounting screws.
3) Replace the cover.

F0803.EPS

Press
forward

Output terminal cable

LCD board
assembly

Integral
indicator

Cover

Boss

CPU assembly

Flat cable

Bracket
(for zero-adjustment
screw pin)

Mounting
screw

Zero-adjustment
screw pin

Figure 8.4.2 Removing and Attaching LCD Board

Assembly and CPU Assembly