8-2

8. MAINTENANCE

Table 8.2.1 Tools for Disassembly and Reassembly

Name

Remarks

Yokogawa-recommended Instrument

Power

supply

Model SDBT or SDBS distributor

Load

resistor

Model 2792 standard resistor [250

±

0.005%, 3 W]

Load adjustment resistor [100

±

1%, 1 W]

Voltmeter

Model 2501 A digital multimeter
Accuracy (10V DC range):

±

(0.002% of rdg + 1 dgt)

Digital

manometer

Model MT110, MT120 precision digital manometer
1) For 10 kPa class

Accuracy:

±

(0.015% of rdg + 0.015% of F.S.)

±

(0.2% of rdg + 0.1% of F.S.)

2) For 130 kPa class

Accuracy:

±

0.02% of rdg

±

5digits

±

(0.2% of rdg + 0.1% of F.S.)

3) For 700 kPa class

Accuracy:

±

(0.02% of rdg + 3digits)

±

5 digits

±

(0.2% of rdg + 0.1% of F.S.)

4) For 3000 kPa class

Accuracy:

±

(0.02% of rdg + 10 digits)

±

(0.2% of rdg + 0.1% of F.S.)

5) For 130 kPa abs class

Accuracy:

±

(0.03% of rdg + 6 digits)

4 to 20 mA DC signal

Select a manometer having
a pressure range close to
that of the transmitter.

Pressure
generator

Model 2657 pneumatic pressure standard for 200 kPa {2 kgf/cm

2

}, 25 kPa {2500 mmH

2

O}

Accuracy:

±

0.05% of F.S. or

±

0.1% setting (whichever is greater)

Requires air pressure
supply.

Pressure

source

Model 6919 pressure regulator (pressure pump)

Pressure range: 0 to 133 kPa {1000 mmHg}

Prepare the vacuum pump
for negative pressure
ranges.

Dead weight gauge tester 25 kPa {2500mmH

2

O}

Accuracy:

±

0.03% of setting

Select a pressure generator
having a pressure range
close to that of the transmitter.

T0801.EPS

for 0 to 10 kPa
for -10 to 0 kPa

for 25 to 130 kPa
for 0 to 25 kPa
for -80 to 0 kPa

for 100 to 700 kPa
for 0 to 100 kPa
for -80 to 0 kPa

for 0 to 3000 kPa
for -80 to 0 kPa

for 0 to 130 kPa abs

. . . . .

. . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . .

. . . . . . . . . . .

. . . . . . . . .

. . . . . . . . . . . .

Note : The above table contains the instruments capable of performing calibration to the 0.2% level. Since special maintenance and management

procedures involving traceability of each instrument to higher-level standards are required for calibration to the 0.1% level, there are
difficulties in calibration to this level in the field.
For calibration to the 0.1% level, contact Yokogawa representatives from which the instrument was purchased or the nearest Yokogawa
office.

P

P

V

Mating calibration flange

Load resistance, 250

Model MT110, MT120
precision digital manometer

High pressure side
diaphragm seal

Rc

R

Load adjustment
resistance, 100

Digital voltmeter

Power
supply

E

Low pressure side diaphragm seal

Model 2657 pneumatic
pressure standards

Pressure source

Supply pressure

If a pressure
generator is used:

If a pressure source
and a manometer
are combined:

F0801.EPS

Figure 8.3.1 Instrument Connections

Advertising