IM 1C22H1-01E

7-11

7. BRAIN TERMINAL BT200 OPERATION

b. Setting User-set Engineering Unit

(D21: DISP UNIT)

This parameter allows entry of the engineering units to
be displayed on the BT200. When the instrument is
shipped, this is set as specified in the order.

Follow the procedure below to change this setting.

Since these units are not displayed on the integral
indicator, use the adhesive labels provided. This
parameter need not be set for % display.

• Example: Set an engineering unit M.

CODE

CAPS

CLR

ESC

SET
D21:DISP UNIT
M

FEED

NO

OK

SET
D21:DISP UNIT

M_

Set “M.”

Press the key twice to

enter the setting.

Press the (OK) key.

F0721.EPS

c. Lower and Higher Range Value Setup

in Engineering Unit (D22: DISP LRV,
D23: DISP HRV)

These parameter items are used to set the lower and
higher range values for the engineering unit display.

When the instrument is shipped, these are set as
specified in the order. Follow the procedure below to
change these settings. Note that these parameters need
not be set for % display.

• Example: Set lower range value (LRV) to –50

and higher range value (HRV) to 50.

FEED

NO

OK

SET
D23:DISP HRV
50M

Press the (OK) key.

F0722.EPS

DEL

CLR

ESC

DATA

DIAG

PRNT

ESC

PARAM
D21:DISP UNT
M
D22:DISP LRV
– 50M
D23:DISP HRV
50M

DEL

CLR

ESC

SET
D22:DISP LRV
0M
- 50

Set “–50.”

Press the key twice to

enter the setting.

Setting LRV

Setting HRV

Set “50.”

Press the key twice to

enter the setting.

SET
D23:DISP HRV
100M
+ 50

(7) Unit Setup for Displayed Temperature

(D30: TEMP UNIT)

When the instrument is shipped, the temperature units
are set to

degC

. Follow the procedure below to

change this setting. Note that changing the unit here
changes the unit for

A20: AMP TEMP

(amplifier

temperature) and

A21: CAPSULE TEMP

(capsule

temperature).

• Example: Change the unit for the temperature

display.

ESC

SET
D30:TEMP UNIT
deg C
< deg C >
< deg F >

Use the or key to

select “deg F.”

Press the key twice to

enter the setting.

F0723.EPS