Physical specifications, Settings when shipped – Yokogawa EJA118 User Manual

Page 55

Advertising
background image

IM 1C22H1-01E

9-2

9. GENERAL SPECIFICATIONS

Physical Specifications

Wetted Parts Materials:

Diaphragm and other wetted parts;

See ‘Model and Suffix Codes’

Non-wetted Parts Materials:

Capillary tube;

SUS316

Protection tube;

SUS304, PVC-sheathed [Max. operating tempera-
ture: 100

°

C (212

°

F) ]

Fill Fluid;

See Table 1.

Housing;

Low copper cast-aluminum alloy with polyurethane
paint (Munsell 0.6GY3.1/2.0)

Enclosure Classification;

JIS C0920 immersion proof (equivalent to NEMA
4X and IEC IP67)

Cover O-rings;

Buna-N

Data plate and tag;

SUS304

Weight:

21.7 kg (47.8 lb): Model EJA118W, 3-inch ANSI
Class 150 flange, without mounting bracket. Add
1.4 kg (3.1 lb) for JIS SCS14A stainless steel
amplifier housing.

Connections:

Refer to the ‘Model and Suffix Codes’ to specify
the process and electrical connection type.

Process temperature
for fill fluid code B

Process temperature
for fill fluid code A

Process temperature
for fill fluid code C

Flange max.
working
pressure

Atmospheric
pressure

Transmitter ambient
temperature range
(For fill fluid code A,B)

Working
pressure

2.7{20}

Process Temperature (

Њ

C)

kPa abs

{mmHg abs}

100{750}

0.1{0.75}

1{7.5}

10{75}

-50

0

50

100 150 200 250 300

F0901.EPS

Figure 1. Working Pressure and Process Temperature

Table 1. Process Temperature and Ambient Temperature

Process

Temperature

(Note 1)

Ambient

temperature

(Note 2)

Working pressure

Specific

gravity

(Note 3)

See Figure 1.

0.94

1.07

1.09

–10 to 250

Њ

C

(14 to 482

Њ

F)

–10 to 60

Њ

C

(14 to 140

Њ

F)

–30 to 180

Њ

C

(–22 to 356

Њ

F)

–15 to 60

Њ

C

(5 to 140

Њ

F)

10 to 300

Њ

C

(50 to 572

Њ

F)

10 to 60

Њ

C

(50 to 140

Њ

F)

–20 to 120

Њ

C

(–4 to 248

Њ

F )

–10 to 60

Њ

C

(–14 to 140

Њ

F )

–50 to 100

Њ

C

(–58 to 212

Њ

F)

–40 to 60

Њ

C

(–40 to 140

Њ

F)

Fill Fluid
Code ‘A’

Fill Fluid
Code ‘B’

Fill Fluid

Code ‘C’

Fill Fluid

Code ‘D’

Fill Fluid
Code ‘E’

51 kPa abs or more

{380 mmHg abs}

Vacuum pressure

not allowed

1.90 to 1.92

1.09

Silicone Oil

Fluorinated Oil

Ethylene Glycol

T0903.EPS

Note 1: See Figure 1. ‘Working Pressure and Process Temperature.’
Note 2: This ambient temperature is the transmitter ambient temperature.
Note 3: Approximate values at a temperature of 25

°

C(77

°

F)

Note 4: The differential pressure transmitter should be installed at least 600 mm below the high

pressure(HP) process connection. However, this value(600 mm) may be affected by ambient
temperature, operating pressure, fill fluid or material of the wetted diaphragm. Contact
YOKOGAWA when the transmitter can not be installed at least 600 mm below the HP process
connection.

<Settings When Shipped

>

Tag Number

Output Mode

Display Mode

Operation Mode

Damping Time

Constant

As specified in order

*1

‘Linear’ unless otherwise specified in order

‘2 sec.’

‘Normal’ unless otherwise specified in order

‘Linear’ unless otherwise specified in order

Calibration Range

Units

As specified in order

As specified in order

Calibration Range

Higher Range Value

Calibration Range

Lower Range Value

Selected from mmH

2

O, mmAq, mmWG,

mmHg, Pa, hPa, kPa, MPa, mbar, bar,
gf/cm

2

, kgf/cm

2

, inH

2

O, inHg, ftH

2

O, or

psi. (Only one unit can be specified)

T0909.EPS

Note 1: If Tag No. is no more than 16 alphanumeric characters (including - and ·), it will be written into

the tag plate and amplifier memory settings.

Advertising