Yokogawa EJA118 User Manual

Page 60

Advertising
background image

IM 1C22H1-01E

9-7

9. GENERAL SPECIFICATIONS

Item

Description

Code

T0908.EPS

Gold-plated diaphragm

T

Teflon film

(Note 3)

With FEP film and fluorinated oil
Working range: 20 to 150

°

C, 0 to 2 MPa {0 to 20 kgf/cm

2

} (Not usable under vacuum)

R

Operating temperature
correction

(Note 4)

Adjusting range: 80 to 300

°

C

V

Capillary without PVC
sheaths

When ambient temperature exceeds 100

°

C, or use of PVC is prohibited

C1

Failure alarm down-scale

(Note 5)

Output status at CPU failure and hardware error: –5%, 3.2 mA DC or less.
When combining with Optional code F1, output signal is –2.5%, 3.6 mA DC or less.

E1

A1

Stainless steel amplifier
housing

(Note 6)

Gold-plate

Amplifier housing material: JIS SCS14A stainless steel (equivalent to JIS SUS316 cast
stainless steel or ASTM CF-8M)

Update time: 0.125 sec or less, see GS for the response time

F1

Fast response

Oil-prohibited use

K5

Degrease cleansing and dehydrating treatment

Oil-prohibited use with
dehydrating treatment

Y

Sealant (liquided silicone rubber)is coated on surfaces of JIS SUS630 nuts used for cover
flange mounting.

Sealing treatment to JIS
SUS630 nuts

Painting

Lightning protector

K1

Degrease cleansing treatment

No serration

(Note 2)

Q

No serration work on the flange gasket surface (for ANSI flange only)

A

Transmitter power supply voltage: 10.5 to 32 V DC (9 to 32 V DC for Fieldbus
communication type) Allowable current: Max. 6000 A (1x40

µ

s),

Repeating 1000 A (1x40

µ

s) 100 times

P

X1

Amplifier cover only

D1

D3

P calibration (psi unit)

bar calibration (bar unit)

(See Table for Span and

Range Limits.)

Calibration units

(Note 1)

D4

M calibration (kgf/cm

2

unit)

Epoxy resin-baked coating

Color change

Coating change

Pressure test /Leak test
Certificate

M05

M06

M07

T31

T32

T33

T36

T37

T38

T39

Mill Certificate

(Test pressure)

2 MPa {20 kgf/cm

2

}

5 MPa {50 kgf/cm

2

}

10 MPa {100 kgf/cm

2

}

3 MPa {29.8 kgf/cm

2

}

7.7 MPa {77 kgf/cm

2

}

7 MPa {70 kgf/cm

2

}

14 MPa {140 kgf/cm

2

}

(Flange rating)

JIS 10K

JIS 20K

JIS 40K

ANSI/JPI Class 150

ANSI/JPI Class 300

ANSI/JPI Class 300

ANSI/JPI Class 600

(For model EJA118W)

(For model EJA118N)

Nitrogen (N

2

) Gas

Retention time: 10 minutes

Process flange, Block

Process flange, Block, Pipe, Base

High pressure side: Process flange, Block, Pipe, Base
Low pressure side: Process flange, Block

(For model EJA118Y)

Note 1: The unit of MWP (Max. working pressure) on the data plate of a housing is same unit as

specified by option code “D1, D3 and D4.”

Note 2: This item cannot be applied to Model EJA118W wetted parts material code H (Hastelloy C), T

(Tantalum), or U (Titanium). (In case for code H, T or U, serration work on the flange gasket
surface is not possible).

Note 3: Teflon film can only be specified for model EJA118W.
Note 4: Specify the process operating temperature for zero correction.

Example: Zero correction by process temperature 90

°

C.

Note 5: The hardware error indicates faulty amplifier or capsule. Standard output status (without /C1) is

up-scale of 110%, 21.6 mA DC or more.

Note 6: Applicable only for electrical connection code ‘2’, ‘3’ or ‘4’. Not applicable for optional code P

and X1. Not applicable for optional code JF1 and JS1.

Advertising