6) integral indicator scale setup, 6) integral indicator scale setup -10, User-set engineering unit display – Yokogawa EJA530A User Manual

Page 41: For % display, set this parameter only

Advertising
background image

IM 01C21F01-01E

8-10

8. BRAIN TERMINAL BT200 OPERATION

(6) Integral Indicator Scale Setup

The following 5 displays are available for integral indicators.

T0808.EPS

D20: DISP SELECT

Display

Related Parameters

NORMAL %

USER SET

USER & %

INP PRES

PRES & %

Indicates –5 to 110%
range depending on the
Measurement range
(C21, C22).

Indicates values depending
on the Engineering range
(D22, D23).

(Note 1)

Units set using Engineering
unit (D21) are not indicated.

Indicates user set and %
alternately in 3 second
intervals.

Indicates input
pressure.

(Note 2)

Indication limits –19999
to 19999.

Indicates input pressure
and % alternately in 3
second intervals.

A10:OUTPUT (%)

45.6 %

Description

A11:ENGR.OUTPUT

20.0 M

A10:OUTPUT (%)

45.6 %

A11:ENGR. OUTPUT

20.0 M

A40:INPUT

456 kPa

A10:OUTPUT (%)

45.6 %

A40:INPUT

456 kPa

(Note 1) Scale range can be specified with range limit specifications up to 5 digits (excluding any

decimal point) for low or high range limits within the range of –19999 to 19999.
The range with decimals is available to the third decimal place.

(Note 2) It indicates the value after zeroing.

See (a.) through (c.) for each setting procedure.

% indication and input
pressure indication

Transmitter is set
for “% display” when
shipped.

D20: DISP SELECT
NORMAL %
INP PRES
PRES & %

User-set engineering unit display

D20: DISP SELECT
USER SET
USER & %

D21: DISP UNIT

D22: DISP LRV

D23: DISP HRV

Set for user-set
engineering unit display.

Set a unit to be
displayed on the BT200.

For % display, set this
parameter only.

Set a numeric value for
engineering unit for 4 mA
output (LRV).

Set a numeric value for
engineering unit for 20 mA
output (HRV).

F0818.EPS