2 replacing the cpu board assembly, Replacing the cpu board assembly -4 – Yokogawa EJA530A User Manual

Page 53

Advertising
background image

IM 01C21F01-01E

9-4

9. MAINTENANCE

᭿ Attaching the Integral Indicator
Integral indicator can be installed in the following three
directions.

F0910.EPS

Figure 9.4.1 Installation Direction of Indicator

1) Align both the LCD board assembly and CPU

assembly connectors and engage them.

2) Insert and tighten the two mounting screws.
3) Replace the cover.

F0903.EPS

Output terminal cable

LCD board
assembly

Integral
indicator

Cover

Boss

CPU assembly

Flat cable

Bracket
(for zero-adjustment
screw pin)

Mounting screw

Zero-adjustment
screw pin

Press

Figure 9.4.2 Removing and Attaching LCD Board Assem-

bly and CPU Assembly

9.4.2 Replacing the CPU Board Assembly

This subsection describes the procedure for replacing
the CPU assembly. (See Figure 9.4.2)

᭿ Removing the CPU Assembly
1) Remove the cover. If an integral indicator is

mounted, refer to Subsection 9.4.1 and remove the
indicator.

2) Turn the zero-adjustment screw to the position

(where the screw head slot is horizontal) as shown
in Figure 9.4.2.

3) Disconnect the output terminal cable (cable with

brown connector at the end). When doing this,
lightly press the side of the CPU assembly connec-
tor and pull the cable connector to disengage.

4) Use a socket driver (width across flats, 5.5mm) to

loosen the two bosses.

5) Carefully pull the CPU assembly straight forward to

remove it.

6) Disconnect the flat cable (cable with black connec-

tor at the end) that connects the CPU assembly and
the capsule.

NOTE

Be careful not to apply excessive force to the
CPU assembly when removing it.

᭿ Mounting the CPU Assembly
1) Connect the flat cable (with black connector)

between the CPU assembly and the capsule.

2) Connect the output terminal cable (with brown

connector).

NOTE

Make certain that the cables are free of pinching
between the case and the CPU assembly edge.

3) Align and engage the zero-adjustment screw pin

with the groove on the bracket on the CPU assem-
bly. Then insert the CPU board assembly straight
onto the post in the amplifier case.

4) Tighten the two bosses. If the transmitter is

equipped with an integral indicator, refer to Subsec-
tion 9.4.1 to mount the indicator.

NOTE

Confirm that the zero-adjustment screw pin is
placed properly in the groove on the bracket
prior to tightening the two bosses. If it is not, the
zero-adjustment mechanism will be damaged.

5) Replace the cover.

Advertising