<7. Wiring>

7-2

IM 01C25R01-01E

(2) Pulse output and Alarm, Status Output
This instruments uses three wires between the

converter and the power supply.
A DC power and load resistance are required, and

pulse output is connected to a totalizer or an electric

counter.
Low level of the pulse output is 0 to 2V. No

communication is possible over a transmission line.

(3) Simultaneous Analog-Pulse Output
When using the simultaneous analog-pulse

output mode, the communicable distance of the

transmission line is restricted on the wiring method.
Table 7.2 shows the examples of connection for this

output mode.

IMPORTANT

For pulse output and the simultaneous analog-

pulse output, use the load resistance. Refer to

Table 7.2.

7.4 Connection

7.4.1 Power Supply Wiring Connection

Table 7.2 shows the wiring example according to

the output types.

7.4.2 External Indicaror Connection

Connect wiring for external indicators to the CHECK

+ and – terminals.

(Note) Use a external indicator whose internal resistance is 10Ω

or less.

Transmitter terminal box

External indicator

F0702.ai

Power supply

+

SUPPL

Y

PULSE

CHECK

ALARM

Figure 7.2

External Indicator Connection

7.4.3 Communicator Connection

Connect the HART Hand Held Terminal (HHT) to

the SUPPLY + and – terminals.

SUPPL

Y

PULSE

CHECK

ALARM

Transmitter terminal box

HART HHT

F0703.ai

Power supply

+

Ignore the polarity

since the HART HHT

is AC-coupled to the

terminal box.

Figure 7.3

HART HHT Connection

7.4.4 Check Meter Connection

Connect the check meter to the CHECK + and

– terminals. (Use hooks.)

• A 4 to 20 mA DC output signal from the CHECK

+ and – terminals.

(Note) Use a check meter whose internal resistance is 10 or less.

SUPPL

Y

PULSE

CHECK

ALARM

Transmitter terminal box

F0704.ai

Power supply

+

Check meter

Figure 7.4

Check Meter Connection

7.4.5 External Temperature Connection

Connect the RTD cable assembly to the Juck

Terminal.

F0705.ai

SUPPL

Y

PULSE

CHECK

ALARM

Figure 7.5

External temperature Connection