3 optional specifications, Optional specifications -10 – Yokogawa EJX930A User Manual

Page 74

Advertising
background image

<10. General Specifications>

10-10

IM 01C25R01-01E

10.3 Optional Specifications

Item

Description

Code

Factory Mutual (FM) FM Explosionproof

*4

Applicable Standard: FM3600, FM3615, FM3810, ANSI/NEMA 250

Explosionproof for Class I, Division 1, Groups B, C and D, Dust-ignitionproof for Class

II/III, Division 1, Groups E, F and G, in Hazardous locations, indoors and outdoors

(NEMA TYPE 4X) “FACTORY SEALED, CONDUIT SEAL NOT REQUIRED.”

Temperature class: T6, Amb. Temp.: –40 to 60°C (–40 to 140°F)

FF1

FM Intrinsically Safe and Nonincendive

*1 *3 *4

Applicable Standard: FM3600, FM3610, FM3611, FM3810, ANSI/NEMA 250,

IEC60079-27

Intrinsically Safe for Class I,II, & III, Division 1, Groups A,B,C,D,F & G, Entity, FISCO.

Class I, Zone 0, AEx ia IIC,

Enclosure: "NEMA TYPE 4X",

Temp. Class: T4, Amb. Temp.: –40 to 60°C (–40 to 140°F).

Intrinsically Apparatus Parameters:

[FISCO (IIC)] Ui=17.5 V, Ii=380 mA, Pi=5.32 W, Ci=3.52 nF, Li=0 μH

[FISCO (IIB)] Ui=17.5 V, Ii=460 mA, Pi=5.32 W, Ci=3.52 nF, Li=0 μH

[Entity] Ui=24 V, Ii=250 mA, Pi=1.2 W, Ci=3.52 nF, Li=0 μH

Sensor Circuit: Uo=6.51 V, Io=4 mA, Po=6 mW, Co=34 μF, Lo=500 mH

Nonincendive for Class I, Division 2, Groups A, B, C and D, NIFW, FNICO

Class I, Zone 2, Group IIC, NIFW, FNICO

Class II, Division 2, Groups F&G, and Class III, Division 1

Enclosure: "NEMA 4X", Temp. Class: T4, Amb. Temp.: –40 to 60°C (–40 to 140°F)

Nonincendive Apparatus Parameters : Vmax.= 32 V, Ci = 1.76 nF, Li = 0 μH

FS15

ATEX

ATEX Flameproof

*4

Applicable Standard: EN 60079-0:2009, EN 60079-1:2007, EN 60079-31:2009

Certificate: KEMA 07ATEX0109 X

II 2G, 2D Ex d IIC T6...T4 Gb, Ex tb IIIC T85°C Db IP6X

Degree of protection: IP66 and IP67

Amb. Temp. (Tamb) for gas-proof: T4; –50 to 75°C (–58 to 167°F),

T5; –50 to 80°C (–58 to 176°F), T6; –50 to 75°C (–58 to 167°F)

Max. process Temp. for gas-proof (Tp): T4; 120°C (248°F), T5; 100°C (212°F),

T6; 85°C (185°F)

Max. surface Temp. for dust-proof: T85°C (Tamb: –30 to 75°C, Tp: 85°C)

*5

KF22

ATEX Intrinsically safe

*2 *3 *4

Applicable Standard: EN 50014, EN 50020, EN 50284, EN 50281-1-1

Certificate: KEMA 06ATEX0037X II 1G,

1D EEx ia IIC T4 Degree of protection: IP66 and IP67

Amb. Temp. (Tamb) for gas-proof: –50 to 60°C (–58 to 140°F)

Maximum Process Temp.(Tp) for gas-proof: 120°C

Electrical data:

[Supply/Output circuit (terminals + and –)] Ui=30 V, Ii=200 mA, Pi=0.9 W, Ci=10 nF,

Li=0 mH

[Pulse Output circuit (terminals – and pulse)] Ui=30 V, Ii=200 mA, Pi=0.9 W, Ci=10 nF,

Li=0 mH

[External Temperature Input circuit (connector)] Uo=30 V, Io=95.4 mA, Po=468 mW,

Co=11 nF, Lo=3.9 mH

Max. surface Temp. for dust-proof: T85°C (Tamb: –40 to 60°C, Tp: 80°C),

T100°C (Tamb: –40 to 60°C, Tp: 100°C), T120°C (Tamb: –40 to 60°C, Tp: 120°C)

*5

KS2

Combined KF22, KS2 and Type n

*1 *4

Type n

Applicable Standard: EN 60079-0, EN 60079-15

II 3G Ex nL IIC T4 Gc, Amb. Temp: –30 to 60 °C

*5

[Supply and Pulse circuit] Ui=30 V DC, Ci=10 nF, Li=0 mH

[Sensor circuit] Uo=7.4 V DC, lo=25 mA, Po=46.3 mW, Co= 11 nF, Lo=3.9 mH

KU22

Advertising
This manual is related to the following products: