<9. Maintenance>

9-2

IM 01C25R01-01E

9.3.2 External Temperature (RTD)

Using a thermometer resistor as input, calibration

of the temperature transmitter is carried out via a

3-core wire connection.
As defined the reference resistor value table of

the thermometer resistor (RTD), obtain resistance

values corresponding to 0, 25, 50, 75 or 100%

of the span, and use the obtained resistance as

the input value, then deliver it to the temperature

transmitter by means of a variable resistor. Measure

the resulting output signal with the voltmeter (digital

multimeter) and check the output value relative to

the input value.

If the output signal deviates from the given range

of accuracy when a given input signal is delivered,

adjust the output using the handheld terminal.

For details of how to adjust the output, refer to

the additional references, “HART Protocol” IM

01C25R02-01E and the instruction manual for each

terminal.

Table 9.1

Instruments Required for Calibration

Name

Yokogawa-recommended Instrument

Remarks

Power supply Model SDBT or SDBS distributor

4 to 20 mA DC signal

Load resistor

Model 2792 standard resistor [250 Ω ±0.005%, 3 W]
Load adjustment resistor [100 Ω ±1%, 1 W]

Voltmeter

Model 2501 A digital multimeter

Accuracy (10V DC range): ±(0.002% of rdg + 1 dgt)

Digital

manometer

Model MT220 precision digital manometer

1) For 10 kPa class

Accuracy: ±(0.015% of rdg + 0.015% of F.S.). . . . for 0 to 10 kPa

±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -10 to 0 kPa

2) For 130 kPa class

Accuracy: ±0.02% of rdg. . . . . . . . . . . . . . . . . . . . for 25 to 130 kPa

±5 digits . . . . . . . . . . . . . . . . . . . . . . . . for 0 to 25 kPa

±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -80 to 0 kPa

3) For 700 kPa class

Accuracy: ±(0.02% of rdg + 3 digits) . . . . . . . . . . . for 100 to 700 kPa

±5 digits . . . . . . . . . . . . . . . . . . . . . . . . for 0 to 100 kPa

±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -80 to 0 kPa

4) For 3000 kPa class

Accuracy: ±(0.02% of rdg + 10 digits). . . . . . . . . . for 0 to 3000 kPa

±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -80 to 0 kPa

5) For 130 kPa abs class

Accuracy: ±(0.03% of rdg + 6 digits) . . . . . . . . . . . for 0 to 130 kPa abs

Select a manometer having

a pressure range close to

that of the transmitter.

Pressure

generator

Model 7674 pneumatic pressure standard for 200 kPa {2 kgf/cm

2

}, 25 kPa

{2500 mmH

2

O}

Accuracy: ±0.05% of F.S.

Requires air pressure

supply.

Dead weight gauge tester 25 kPa {2500 mmH

2

O}

Accuracy: ±0.03% of setting

Select the one having a

pressure range close to that

of the transmitter.

Pressure

source

Model 6919 pressure regulator (pressure pump)

Pressure range: 0 to 133 kPa {1000 mmHg}

Prepare the vacuum pump

for negative pressure

ranges.

Variable

resistor

279301 type 6-dial variable resistor (accuracy: ±0.005%)

For calibration of

thermometer resistor (RTD)

input

Note: The above table contains the instruments capable of performing calibration to the 0.2% level. Since special maintenance and

management procedures involving traceability of each instrument to higher-level standards are required for calibration to the 0.1%

or higher level, there may be difficulties in calibration to this level in the field. For calibration to the 0.1% level, contact Yokogawa

representatives from which the instrument was purchased or the nearest Yokogawa office.