<2. About the EJX Multivariable Transmitter>

2-1

IM 01C25R01-01E

2. About the EJX Multivariable

Transmitter

This chapter gives an overview of the functions and

the installation of the EJX Multivariable transmitter.

For details on specific procedures, refer to the

corresponding chapter.

2.1 Features

• Multi Sensing Function

The EJX multivariable transmitter has a Multi

Sensing function that enables a single transmitter

to measure differential pressure, static pressure,

and external temperature. Mass flow measurement

instruments can calculate and output the flow

value using these three measured variables except

Modbus Communication Type. The FSA120 flow

configuration software (FlowNavigator) is used to

configure mass flow calculation.

• Simultaneous Analog Output and Pulse

Output (HART protocol type)

The EJX multivariable transmitter has one analog

output for the output of single measured variables.

With digital communications, all of these variables

can be output simultaneously.
In addition to analog output, pulse output is

provided as a standard function of the EJX

multivariable transmitter. It can perform both types

of output simultaneously. Pulse output is used for

total flow, flow rate, and alarm status.

DP sensor

SP sensor

Sensor Capsule

A/D

ROM

PC: Flow configuration

Electronics module

RTD Temperature Input

Pressure Input

4 to 20 mA output

for selected PV.

RAM

Differential Pressure

/Static Pressure

/Process Temperature

/Calculated Mass Flow

ROM

MPU

/Calculate Mass Flow

/System control

D/A

Digital I/O

Pulse

output

Pulse output

or Status output.

Silicon Resonant

sensor

F0201.ai

Figure 2.1

Block Diagram of EJX Multivariable Transmitter System (HART protocol type)

2.2 Initial Check and Installation

Procedure

Figure 2.2 is a flowchart showing the basic

sequence for installing and wiring an EJX

Multivariable Transmitter.
Refer to the specified chapter for the details of each

procedure.
Flow configuration can be performed at the

following stages:
(1) On the bench, before installation.
(2) In the field, after installation.

IMPORTANT

Before performing flow configuration in the field,

make sure the installation location meets all

safety requirements.