() Technical Data

a) ATEX Intrinsically Safe “ia”
Caution for ATEX Intrinsically safe “ia”

Note 1. Model YTA110/KU2, YTA310/KU2 and

YTA320/KU2 temperature transmitters for

potentially explosive atmospheres:

• No. KEMA 02ATEX1026X

• Applicable Standard: EN 60079-0:2009,

EN 60079-11:2007, EN 60079-26:2007,

EN 60529:1991

• Type of Protection and Marking code: II 1 G Ex

ia IIC T4...T5

• Temperature Class: T5, T4

• Ambient Temperature: –40 to 70°C for T4,

–40 to 50°C for T5

• Enclosure: IP67

Note 2. Electrical Data

• In type of explosion protection intrinsic safety II

1 G Ex ia IIC only for connection to a certified

intrinsically safe circuit with following maximum

values:

• [Supply circuit]

Ui = 30 V

Ii = 165 mA

Pi = 900 mW

Effective internal capacitance, Ci = 20 nF

Effective internal inductance, Li = 730 μH

• [Sensor circuit]

Uo = 9 V

Io = 40 mA

Po = 90 mW

Max. allowed external capacitance, Co = 0.7μF

Max. allowed external inductance, Lo = 10 mH

• The above parameters apply when one of the

two conditions below is given:

- the total Li of the external circuit (excluding

the cable) is < 1% of the Lo value or

- the total Ci of the external circuit (excluding

the cable) is < 1% of the Co value.

• The above parameters are reduced to 50%

when both of the two conditions below are

given:

- the total Li of the external circuit (excluding

the cable) is ≥ 1% of the Lo value and

- the total Ci of the external circuit (excluding

the cable) is ≥ 1% of the Co value.

• The reduced capacitance of the external circuit

(including cable) shall not be greater than 1μF

for Group IIB and 600nF for Group IIC.

Note 3. Installation

• All wiring shall comply with local installation

requirements. (Refer to the installation diagram)

Note 4. Operation

• Keep strictly the “WARNING” on the label on

the transmitter.

WARNING: POTENTIAL ELECTROSTATIC

CHARGING HAZARD. SEE

USER’S MANUAL BEFORE USE.

Note 5. Maintenance and Repair

• The instrument modification or parts

replacement by other than authorized

representative of Yokogawa Electric

Corporation is prohibited and will void ATEX

Intrinsically safe “ia” Certification.

Note 6. Specific condition of use

• Because the enclosure of the Temperature

Transmitter is made of aluminium, if it is

mounted in an area where the use of category

1G apparatus is required, it must be installed

such, that, even in the event of rare incidents,

ignition source due to impact and friction sparks

are excluded.

• Avoid any actions that cause the generation of

electrostatic charge on the non-metallic parts,

such as rubbing with a dry cloth on coating face

of product.

Transmitter

Supply

Safety Barrier *1

Hazardous Location

Nonhazardous Location

+

+

[Installation Diagram]

F0205.ai

Sensor

1
2
3
4
5

*1: In any safety barriers used the output current must be

limited by

a resistor “R” such that Imaxout-Uz/R.

b) ATEX Flameproof Type and Dust Ignition

Proof Type

Caution for ATEX Flameproof Type and Dust

Ignition Proof Type

Note 1. Model YTA110/KU2, YTA310/KU2 and

YTA320/KU2 temperature transmitters are

applicable for use in hazardous locations:

• No. KEMA 07ATEX0130

• Applicable Standard: EN 60079-0:2006,

IEC 60079-1:2007, EN 61241-0:2006,

EN 61241-1:2004

• Type of Protection and Marking Code: II 2 G

Ex d IIC T6/T5, II 2 D Ex tD A21 IP67 T70°C,

T90°C

• Ambient Temperature for Gas Atmospheres:

–40 to 75°C (T6), –40 to 80°C (T5)

Advertising