Electrical data are as follows;

Supply Input (Terminals: + and -)

Ui = 30 V

Ci = 28 nF

Li = 730 μH

Sensor Output (Terminals: 1 to 5)

Uo = 9 V

Io = 40 mA

Po = 90 mW

Co = 0.7 μF

Lo = 10 mH

SUPPLY

SUPPLY

SENSOR

Associated Apparatus

Hazardous Area

Non-hazardous Area

+

+

C

F0206.ai

1
2
3
4
5

SENSOR

(2) Electrical Connection
The type of electrical connection is stamped near

the electrical connection port according to the

following marking.

Marking

Screw Size

ISO M20×1.5 female

M
A

ANSI 1/2 NPT female

T0201.ai

F0207.ai

Location of the marking

(3)

Installation

WARNING

All wiring shall comply with local installation

requirement and local electrical code.

(4) Operation

WARNING

• OPEN CIRCUIT BEFORE REMOVING

COVER. INSTALL IN ACCORDANCE WITH

THIS USER’S MANUAL

• Take care not to generate mechanical

sparking when access to the instrument and

peripheral devices in hazardous locations.

(5) Maintenance and Repair

WARNING

The instrument modification or parts replacement

by other than authorized Representative of

Yokogawa Electric Corporation is prohibited and

will void the certification.

(6) Name Plate

F0208.ai

TEMPERATURE

TRANSMITTER

YTA

MODEL

SUFFIX

STYLE

SUPPLY

NO.

OUTPUT

CAL

RNG

*3

MODEL: Specified model code.

SUFFIX: Specified suffix code.

STYLE: Style code.

SUPPLY: Supply voltage.

NO.: Serial number and year of production*

1

.

OUTPUT: Output signal.

FACTORY CAL: Specified calibration range.

TOKYO 180-8750 JAPAN:

The manufacturer name and the address*

2

.

*1: The third figure from the left shows the production year.

The relationship between the production year and the

third figure is shown below.

The third figure

F

G

H

J

K

L

M

The year of

Production

2006 2007 2008 2009 2010 2011 2012