<6. Maintenance>

6-

IM 01C50B01-01E

6. Maintenance

6. General

Each component of this instrument is configured in

units to make maintenance easier.
This chapter contains disassembly and assembly

procedures associated with calibration, adjustment

and part replacement required for maintenance of

the affected instrument.

IMPORTANT

1. Maintenance of this instrument should be

performed in a service shop where the

necessary tools are provided.

2. Handling the CPU assembly

Some of the parts contained in the CPU

assembly are susceptible to static electricity

damage. Before performing maintenance,

use a ground wrist band or other antistatic

mea sures, and avoid touching the electronic

components and circuits with bare hands.

When removed from the instrument, keep

the CPU assembly in an antistatic bag.

6.2 Calibration

This instrument is fully factory-tested and is

guaranteed for the intended accuracy, eliminating

the need for calibration. When calibration needs to

be varified, the following equipment and calibration

procedure is recommended.

6.2. Selection of Equipment for

Calibration

Table 6.1 lists the equipment required for

calibration. The calibration equipment traceable to a

verifying agency standard should be used.

6.2.2 Calibration Procedure

To conduct calibration required to evaluate the

uncertainty while using the instrument, follow the

steps below:
1. In accordance with the example wiring shown in

Figure 6.1, connect each equipment and initiate

warm up. Lay wiring on the input side according

to the sensor to be used.

Table 6.

Calibration Equipment List

Name

Recommended

Remark

Power

supply

SDBT, SDBS type

distributor

4 to 20mA DC

(Output voltage:

26.5±1.5V, drop

by internal 250Ω

resistance included)

Load

resistance

2792 type standard

resistor

(250Ω ±0.005%)

For 4 to 20mA DC

Voltmeter Model 1271

digital multimeter

(accuracy: ±0.002%)

For 4 to 20mA DC

signal

Universal

calibrator

Model 9100 type

For calibration of

DC voltage and

thermocouple

Variable

resistor

279301 type 6-dial

variable resistor

(accuracy: ±0.001%

±2mΩ)

For calibration

of thermometer

resistor (RTD) input

F0601.ai

1

2

3

4

5

1

2

3

4

5

(A)
(B)
(B)

(A)

a. Wiring of power supply and output

b. Example of wiring for thermocouple or DC voltage input
(when 1 input type is used)

+ Output signal

Load

resistance

DC voltage generator

Voltmeter

c. Example of wiring for thermometer resistor 4-core type
(when 1 input type is used)

Variable resistor

(+)

(–)

Figure 6.

Example of Wiring for Calibration

Equipment

2. For DC voltage input

With a voltage generator, deliver input signals

corresponding to 0, 25, 75, or 100% of the input

span to the temperature transmitter. Measure

the resulting input signal with the voltmeter

(digital multimeter) and check the output value

relative to the input value.