Yokogawa YTA320 User Manual

Page 37

Advertising
background image

<7. General Specifications>

7-3

IM 01C50B01-01E

Table 7.

Input Type, Measurement Range and Accuracy

Sensor Type Reference

Standard

Measurement Range

Minimum

Span

(Recommended)

Accuracy

Input range

A/D Accuracy

D/A

Accuracy

°C

°F

°C

°F

°C

°F

T/C

B

IEC584

100 to 1820 212 to 3308

25°C

(45°F)

100

300

400

to

to

to

300

400

1820

212

572

752

to

to

to

572

752

3308

±3.0

±1.0

±0.75

±5.4

±1.8

±1.35

±0.02%

of span

E

-200 to 1000 -328 to 1832

-200

-50

to

to

-50

1000

-328

-58

to

to

-58

1832

±0.35

±0.16

±0.63

±0.29

J

-200 to 1200 -328 to 2192

-200

-50

to

to

50

1200

-328

-58

to

to

-58

2192

±0.40

±0.20

±0.72

±0.36

K

-200 to 1372 -328 to 2502

-200

-50

to

to

-50

1372

-328

-58

to

to

-58

2502

±0.50

±0.25

±0.90

±0.45

N

-200 to 1300 -328 to 2372

-200

-50

to

to

-50

1300

-328

-58

to

to

-58

2372

±0.80

±0.35

±1.44

±0.63

R

-50 to 1768

-58 to 3214

-50

0

100

600

to

to

to

to

0

100

600

1768

-58

32

212

1112

to

to

to

to

32

212

1112

3214

±1.0

±0.80

±0.60

±0.40

±1.8

±1.44

±1.08

±0.72

S

-50 to 1768

-58 to 3214

-50

0

100

600

to

to

to

to

0

100

600

1768

-58

32

212

1112

to

to

to

to

32

212

1112

3214

±1.0

±0.80

±0.60

±0.40

±1.8

±1.44

±1.08

±0.72

T

-200 to 400 -328 to 752

-200

-50

to

to

-50

400

-328

-58

to

to

-58

752

±0.25

±0.14

±0.45

±0.25

W3

ASTM

E988

0 to 2300

32 to 4172

0

400

1400

2000

to

to

to

to

400

1400

2000

2300

32

752

2552

3632

to

to

to

to

752

2552

3632

4172

±0.80

±0.50

±0.60

±0.90

±1.44

±0.90

±1.08

±1.62

W5

0 to 2300

32 to 4172

0

400

1400

2000

to

to

to

to

400

1400

2000

2300

32

752

2552

3632

to

to

to

to

752

2552

3632

4172

±0.70

±0.50

±0.70

±0.90

±1.26

±0.90

±1.26

±1.62

L

DIN43710

-200 to 900 -328 to 1652

-200

-50

to

to

-50

900

-328

-58

to

to

-58

1652

±0.30

±0.20

±0.54

±0.36

U

-200 to 600 -328 to 1112

-200

-50

to

to

-50

600

-328

-58

to

to

-58

1112

±0.50

±0.25

±0.90

±0.45

RTD

Pt100

Pt200

Pt500

JPt100

IEC751

-200

-200

-200

-200

to

to

to

to

850

850

850

500

-328

-328

-328

-328

to

to

to

to

1562

1562

1562

932

10°C

(18°F)

-200

-200

-200

-200

to

to

to

to

850

850

850

500

-328

-328

-328

-328

to

to

to

to

1562

1562

1562

932

±0.14

±0.30

±0.20

±0.16

±0.25

±0.54

±0.36

±0.29

JIS C1604

Cu

SAMA

RC21-4

-70 to 150

-94 to 302

-70

-40

to

to

-40

150

-94

-40

to

to

-40

302

±1.35

±1.0

±2.43

±1.8

Ni120

-70 to 320

-94 to 608

-70 to 320

-94 to 608 ±0.11 ±0.19

mV

-10 to 100 [mV]

3 [mV]

±12 [μV]

ohm

0 to 2000 [Ω]

20 [Ω]

±0.35 [Ω]

Table 7.2

YTA0 Effect of Ambient Temperature

Sensor Type

Temperature Coefficient

Thermocouples E, J, K, N, T, L, U

0.08°C + 0.02% of abs.reading

Thermocouples R, S, W3, W5

0.25°C + 0.02% of abs.reading

T/C B

100°C ≤ Reading < 300°C 1°C + 0.02% of abs.reading
300°C ≤ Reading

0.5°C + 0.02% of abs.reading

RTD

0.08°C + 0.02% of abs.reading

mV

0.002 mV + 0.02% of abs.reading

ohm

0.1Ω + 0.02% of abs.reading

Note1: Ambient Temperature Effect per 10°C change is

±0.1% or ±(temperature coefficient/span), whichever

is greater.

Note2: The “abs.reading” on Table7.2 means the absolute

value of the reading in °C.

Example of abs reading;

When the temperature value is 250 Kelvin, abs reading

is 23.15, absolute (250–273.15).

Example of Ambient Temperature Effect;

Conditions;

1) Input Sensor: Pt100

2) Calibration Range: –100 to 100°C

3) Reading value: –50°C

Ambient Temperature Effect per 10°C;

Temperature Coefficient/Span

=(0.08°C+0.02/100×|–50°C|)/{100°C–(–100°C)}

= 0.00045→ 0.045%

Therefore, Ambient Temperature Effect is ±0.1%/10°C

Advertising