<7. General Specifications>

7-

IM 01C50B01-01E

7. General Specifications

7.1 Standard Specifications

For the specifications of Fieldbus communication

type marked with (◊), refer to IM 01C50T02-01E.

Input signal:

Input points: YTA110; 1, YTA310; 1, YTA320; 2

Thermocouple (T/C):

J, K, E, T, R, S, B, N (JIS/IEC584), L, U

(DIN43710), W3, W5 (ASTM988)

Thermometer resistor (RTD):

Pt100, Pt200, Pt500 (IEC751), JPt100(JIS),

Cu, Ni120 2/3/4-wire

DC voltage: mV

Resistance: Resistor 2/3-wire

Input signal source resistance:

1kΩ or lower (at thermocouple, DC voltage

input)

Input lead wire resistance:

10Ω/1-core or less (with thermometer resistor,

resistance input)

Output(◊):

Two wire 4 to 20 mA DC

Output range: 3.68 mA to 20.8 mA

Measuring range:

Refer to Table 7.1.

Power supply voltage(◊):

For general use type;

10.5 to 42 VDC

With lightning protection (Optional code A);

10.5 to 32 VDC

For intrinsically safe, Type n, non-incendive or

non-sparking type;

10.5 to 30 VDC

Communication line conditions(◊):

Power supply voltage; 16.4 to 42 VDC

Load resistance; 250 to 600Ω (including cable

resistance)

For relation between power supply voltage and

load resistance, refer to Figure 5.3.

Maximum line length; 2 km when CEV cable is

used.

Load resistance; 0.22 μF or less

Load inductance; 3.3 mH or less

Spacing to power line;

Output signal line; 15 cm or longer (avoid

parallel wiring)

Input signal line; 100 cm or longer (avoid

parallel wiring)

Input impedance of receiver connected to

receiver resistance; 10 kΩ or higher (at 2.4 kHz)

Accuracy(◊):

YTA110;

A/C conversion accuracy + D/A conversion

accuracy (Refer to Table 7.1) or ±0.1%,

whichever is greater. For thermocouple input,

above plus reference junction compensation

accuracy.

YTA310, YTA320;

A/C conversion accuracy + D/A conversion

accuracy (Refer to Table 7.1) For

thermocouple input, above plus reference

junction compensation accuracy.

Reference junction compensation accuracy:

±0.5°C (±0.9°F)

Ambient temperature:

–40 to 85°C (–40 to 185°F) . . . .general type

–30 to 80°C (–1 to 176°F) . . . . .with Integral

indicator

Ambient temperature effect per 10°C change(◊):

YTA110;

±0.1% or ± (Temperature Coefficient/span),

whichever is greater (refer to Table 7.2)

YTA310, YTA320;

A/D conversion temperature factor + D/A

conversion temperature factor

(refer to Table 7.3)

Stability

RTD:

±0.1% of reading or ±0.1°C per 2 years,

whichever is greater at 23±2°C.

T/C:

±0.1% of reading or ±0.1°C per year,

whichever is greater at 23±2°C.

5 Year Stability

RTD:

±0.2% of reading or ±0.2°C,

whichever is greater at 23±2°C.

T/C:

±0.4% of reading or ±0.4°C,

whichever is greater at 23±2°C.

Ambient humidity:

5 to 100%RH at 40°C (104°F)