2 withstand voltage test procedure, 1 csa certification, Withstand voltage test procedure -3 – Yokogawa YTA320 User Manual

Page 9: Csa certification -3, Caution

Advertising
background image

<2. Notes on Handling>

2-3

IM 01C50B01-01E

or more is confirmed (or 20MΩ if the unit is

equipped with a built-in arrester).

3. Upon completion of the test, remove the

insulation resistance meter, connect a 100KΩ

resistor between the transition wiring and the

grounding terminal, and allow the electricity to

discharge. Do not touch the terminal with your

bare hands while the electricity is discharging

for more than 1 second.

2.6.2 Withstand voltage test procedure

Testing between the output terminal and

the input terminal

1. Lay transition wiring between the + terminal,

the – terminal, and the check terminal of the

terminal box.

2. Lay transition wiring between terminals 1, 2, 3,

4 and 5 of the terminal box.

3. Connect the withstand voltage tester (with

the power turned OFF) between the transition

wiring shown in Steps 1 and 2 above.

4. After setting the current limit value of the

withstand voltage tester to 10mA, turn

the power ON, and carefully increase the

impressed voltage from 0V to the specified

value.

5. The voltage at the specified value must remain

for a duration of one minute.

6. Upon completion of the test, carefully reduce

the voltage so that no voltage surge occurs.

Testing between the output terminal and

the grounding terminal

1. Lay the transition wiring between the + terminal,

the - terminal and the check terminal of the

terminal box, and connect the withstand voltage

tester (with the power turned OFF) between the

transition wiring and the grounding terminal.

Connect the grounding side of the withstand

voltage tester to the grounding terminal.

2. After setting the current limit value of the

withstand voltage tester to 10mA, turn the

power ON, and gradually increase the

impressed voltage from 0V to the specified

value.

3. The voltage at the specified value must remain

for a duration of one minute.

4. Upon completion of the test, carefully reduce

the voltage so that no voltage surge occurs.

Testing between the input terminal and

the grounding terminal

1. Lay the transition wiring across terminals 1, 2,

3, 4, and 5 of the terminal box and connect the

withstand voltage tester (with the power turned

OFF) between the transition wiring and the

grounding terminal. Connect the grounding side

of the withstand voltage tester to the grounding

terminal.

2. After setting the current limit value of the

withstand voltage tester to 10mA, turn the

power ON, and gradually increase the

impressed voltage from 0V to the specified

value.

3. The voltage at the specified value must remain

for a duration of one minute.

4. Upon completion of the test, carefully reduce

the voltage so that no voltage surge occurs.

2.7 Installation of Explosion

Protected Type Transmitters

In this section, further requirements and differences

and for explosionproof type instrument are

described. For explosionproof type instrument,

the description in this chapter is prior to other

description in this users manual.

CAUTION

To preserve the safety of explosionproof

equipment requires great care during mounting,

wiring, and piping. Safety requirements also

place restrictions on maintenance and repair

activities. Please read the following sections very

carefully.

2.7.1 CSA Certification

Model YTA110/CU1, YTA310/CU1 and YTA320/

CU1 temperature transmitters can be selected the

type of protection (CSA Intrinsically Safe, Non-

incendive, or Explosionproof) for use in hazardous

locations.

Note 1. For the installation of this transmitter, once

a particular type of protection is selected, any

other type of protection cannot be used. The

installation must be in accordance with the

description about the type of protection in this

instruction manual.

Note 2. In order to avoid confusion, unnecessary

marking is crossed out on the label other

than the selected type of protection when the

transmitter is installed.