4.1.3. Procedure "AUTO", "R+TIMER", "R-TIMER"
1. Select the desired mode by means of the

FUNC key.

2. Connect the black and blue cables to the instrument input terminals

B1 and B4

respectively.

B2

B3

B4

B1

Connection of the instrument terminals during LOW

test.


3. If the cables supplied with the instrument are not long enough for the measurement you can

extend the black cable.

4. Connect two crocodiles to the cable terminals.
5. Short-circuit the measuring cable ends making sure that the conductive parts of the

crocodiles make a good contact to each other. Press the

START/STOP key. If the

instrument displays a resistance value other than 0.00 repeat the instrument
calibration
(see paragraph 4.1.1).

6. Connect the instrument terminals to the ends of the conductor on which the continuity test is

to be carried out (see previous picture).

7.

If the mode "R+TIMER" or "R-TIMER" was selected use the following keys to select
the duration time of the test:

press this key to increase the duration time of the test (

Tmax=15 seconds).


press this key to decrease the duration time of the test (

Tmin=3 seconds).



8. Press

the

START/STOP key. The instrument effects the measurement.

In R+/R- Timer mode press

START/STOP key again if the test is to be stopped.


CAUTION

The display of

“Measuring” means that the instrument is measuring. During

this phase never disconnect test leads.