O 5.75, O> 99.9 – HT instruments MACROTEST 5035 User Manual

Page 14

Advertising
background image

MACROTEST

5035

EN - 13

4.1.3.1. Mode

"AUTO"

At the end of the test, if the

average

resistance value Ravg results to be
lower than 5

Ω the instrument emits a

double sound signal indicating the
positive outcome of the test and
displays one screen similar to the
screen alongside.

CAL

LOW

1.07

219

mA

AUTO


SAVING:

The test can be stored pressing the

SAVE key twice

(according to paragraph 5.1).



4.1.3.2. Mode "R+TIMER" or “R-TIMER"

If during the test a

resistance value

R+Timer or R–Timer lower than 5

was detected, the instrument (after the
set time has elapsed) emits a

double

sound signal indicating the positive
outcome
of the test displays one screen
similar to the screen alongside.

CAL LOW

1.08

219

mA

5

s

R+
R-

TIMER


SAVING:

The test can be stored pressing the

SAVE key twice

(according to paragraph 5.1).


4.1.4. Anomalous cases during "AUTO", "R+TIMER", "R-TIMER" tests

In case a value of

Ravg or R+ or R-

higher than or equal to 5

but lower

than 99.9

was detected, at the end of

the test the instrument

emits a long

sound signal and displays one screen
similar to the screen alongside.

CAL LOW

o 5.75

216

mA

5

s


AUTO

In AUTO mode, if a

Ravg or R+ or R-

higher than 99.9

was detected, at

the end of the test the instrument emits
a long sound signal and displays the
screen alongside.

CAL LOW

o> 99.9

- - -

mA


AUTO

Average test
current value
Iavg.

Average
resistance value
Ravg.

Value of the test
current

I+ or I-.

ATTENTION: value of R

avg

higher than 5

99.9

is the

maximum value
which can be
measured in the
LOW

AUTO or

R+ or R-mode.

The symbols

R

+

or

R

-

are displayed.

AUTO mode.

AUTO mode.

ATTENTION: value of R

avg

or R+ or R-

is too

high.

Duration time of
the test.

Only in case the
mode R+TIMER
or R-TIMER was
selected.

SAVE

SAVE